Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.
Peaks | tR (min) | Ion (Quantifier) | Ion (Qualifier) | Dwell Time (ms) | |
---|---|---|---|---|---|
1. | Furan-d4 | 2.428 | 72 | 42 | 50 |
2. | Furan | 2.447 | 68 | 39 | 50 |
3. | 2-Methylfuran-d6 | 4.464 | 88 | 58 | 30 |
4. | 2-Methylfuran | 4.536 | 82 | 53 | 30 |
5. | 3-Methylfuran | 4.846 | 82 | 53 | 30 |
Peaks | tR (min) | Ion (Quantifier) | Ion (Qualifier) | Dwell Time (ms) | |
---|---|---|---|---|---|
6. | 2-Ethylfuran-d5 | 7.001 | 101 | 55 | 30 |
7. | 2-Ethylfuran | 7.100 | 81 | 96 | 30 |
8. | 2,5-Dimethylfuran-d3 | 7.179 | 99 | 84 | 30 |
9. | 2,5-Dimethylfuran | 7.243 | 95 | 67 | 30 |
10. | 2-Pentylfuran-d11 | 11.501 | 149 | 83 | 30 |
11. | 2-Pentylfuran | 11.570 | 138 | 81 | 30 |
Column | Rxi-624Sil MS, 30 m, 0.25 mm ID, 1.40 µm (cat.# 13868) |
---|---|
Standard/Sample | |
Diluent: | N/A |
Conc.: | 10 ng/mL |
Injection | split (split ratio 10:1) |
Liner: | Topaz 1.8 mm ID straight/SPME inlet liner (cat.# 23280) |
Inj. Temp.: | 280 °C |
Split Vent Flow Rate: | 14 mL/min |
Oven | |
Oven Temp.: | 35 °C (hold 3 min) to 75 °C at 8 °C/min to 200 °C at 25 °C/min (hold 1 min) |
Carrier Gas | He, constant flow |
Flow Rate: | 1.4 mL/min |
Detector | MS |
---|---|
Transfer Line Temp.: | 280 °C |
Analyzer Type: | Quadrupole |
Source Temp.: | 325 °C |
Quad Temp.: | 200 °C |
Electron Energy: | 70 eV |
Tune Type: | PFTBA |
Ionization Mode: | EI |
Instrument | Agilent 7890B GC & 5977B MSD |
Sample Preparation | Data was collected by extracting via HS-SPME from a 20 mL vial (cat.# 23083) capped with a magnetic screw-thread cap (cat.# 23091). The vial contained 10 mL of sodium chloride solution spiked with 100 ng of each analyte and internal standard. A Restek PAL SPME Arrow (120 µm carbon wide range [WR]/PDMS; cat.# 27487) was used. SPME Arrow sampling conditions: 10 min extraction time, 10 min incubation at 50 °C, 1 min desorption at 280 °C, agitation at 250 rpm. |
We have determined your location as:
Or select your location below:
Sorry! We are unable to determine your location. Please select your location below.