Peaks | tR (min) | Conc. (µg/mL) | Mix | |
---|---|---|---|---|
1. | Dichlorvos | 4.63 | 60 | 32591 |
2. | Acephate | 5.52 | 60 | 32591 |
3. | Oxamyl | 5.97 | 60 | 32588 |
4. | Flonicamid | 6.21 | 60 | 32588 |
5. | Propoxur | 6.29 | 60 | 32589 |
6. | Ethoprophos | 6.39 | 60 | 32591 |
7. | Naled | 6.53 | 60 | 32591 |
8. | Dimethoate | 6.86 | 60 | 32591 |
9. | Carbofuran | 6.87 | 60 | 32589 |
10. | Diazinon | 7.12 | 60 | 32591 |
11. | Spiroxamine 1 | 7.66 | 60 | 32587 |
12. | Methyl parathion | 7.71 | 60 | 32591 |
13. | Carbaryl | 7.79 | 60 | 32589 |
14. | Metalaxyl | 7.83 | 60 | 32587 |
15. | Spiroxamine 2 | 7.98 | 60 | 32587 |
16. | Methiocarb | 8.02 | 60 | 32588 |
17. | Malathion | 8.10 | 60 | 32591 |
18. | Chlorpyrifos | 8.29 | 60 | 32591 |
19. | MGK-264 1 | 8.52 | 60 | 32587 |
20. | Thiamethoxam | 8.58 | 60 | 32588 |
21. | MGK-264 2 | 8.68 | 60 | 32587 |
22. | Fipronil | 8.84 | 60 | 32588 |
23. | Prallethrin | 9.02 | 60 | 32590 |
24. | Paclobutrazol | 9.28 | 60 | 32589 |
25. | Imidacloprid | 9.40 | 60 | 32586 |
26. | Imazalil | 9.65 | 60 | 32589 |
27. | Fludioxonil | 9.69 | 60 | 32587 |
28. | Myclobutanil | 9.88 | 60 | 32589 |
29. | Cinerin I | 9.88 | 60 | 32590 |
30. | Kresoxim-methyl | 9.94 | 60 | 32587 |
31. | Chlorfenapyr | 10.19 | 60 | 32587 |
Peaks | tR (min) | Conc. (µg/mL) | Mix | |
---|---|---|---|---|
32. | Jasmolin I | 10.58 | 60 | 32590 |
33. | Pyrethrin I | 10.61 | 60 | 32590 |
34. | Pyrethrin I | 10.83 | 60 | 32590 |
35. | Trifloxystrobin | 11.14 | 60 | 32587 |
36. | Propiconazole 1 | 11.14 | 60 | 32589 |
37. | Propiconazole 2 | 11.28 | 60 | 32589 |
38. | Tebuconazole | 11.52 | 60 | 32589 |
39. | Piperonyl butoxide | 11.62 | 60 | 32587 |
40. | Spiromesifen | 12.00 | 60 | 32587 |
41. | Fenoxycarb | 12.16 | 60 | 32589 |
42. | Acetamiprid | 12.19 | 60 | 32587 |
43. | Phosmet | 12.20 | 60 | 32591 |
44. | Bifenthrin | 12.21 | 60 | 32590 |
45. | Bifenazate | 12.25 | 60 | 32587 |
46. | Etoxazole | 12.40 | 60 | 32587 |
47. | Chlorantraniliprole fragment | 12.47 | 60 | 32589 |
48. | Cinerin II | 12.84 | 60 | 32590 |
49. | Jasmolin II | 13.49 | 60 | 32590 |
50. | Pyrethrin II | 13.51 | 60 | 32590 |
51. | cis-Permethrin | 14.03 | 60 | 32590 |
52. | trans-Permethrin | 14.16 | 60 | 32590 |
53. | Pyridaben | 14.20 | 60 | 32588 |
54. | Spirotetramat | 14.65 | 60 | 32586 |
55. | Cyfluthrin | 14.65-14.85 | 60 | 32590 |
56. | Cypermethrin | 14.97-15.18 | 60 | 32590 |
57. | Boscalid | 15.07 | 60 | 32587 |
58. | Acequinocyl | 15.28 | 60 | 32586 |
59. | Etofenprox | 15.28 | 60 | 32590 |
60. | Thiacloprid | 15.73 | 60 | 32588 |
61. | Azoxystrobin | 16.93 | 60 | 32587 |
62. | Chlorantraniliprole fragment | 18.45 | 60 | 32589 |
Column | Rxi-5ms, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13423) |
---|---|
Standard/Sample | Oregon pesticides standard #1 (cat.# 32586) |
Oregon pesticides standard #2 (cat.# 32587) | |
Oregon pesticides standard #3 (cat.# 32588) | |
Oregon pesticides standard #4 (cat.# 32589) | |
Oregon pesticides standard #5 (cat.# 32590) | |
Oregon pesticides standard #6 (cat.# 32591) | |
Diluent: | Acetonitrile |
Conc.: | 60 µg/mL |
Injection | |
Inj. Vol.: | 1 µL splitless (hold 0.5 min) |
Liner: | Topaz 4.0 mm ID single taper inlet liner w/wool (cat.# 23447) |
Inj. Temp.: | 250 °C |
Purge Flow: | 5 mL/min |
Oven | |
Oven Temp.: | 70 °C (hold 1 min) to 220 °C at 30 °C/min to 240 °C at 5 °C/min to 315 °C at 10 °C/min (hold 10 min) |
Carrier Gas | He, constant flow |
Flow Rate: | 1.4 mL/min |
Detector | Thermo TSQ8000 | ||||||||
---|---|---|---|---|---|---|---|---|---|
Scan Program: | |||||||||
| |||||||||
Transfer Line Temp.: | 290 °C | ||||||||
Analyzer Type: | Quadrupole | ||||||||
Source Temp.: | 330 °C | ||||||||
Solvent Delay Time: | 2.5 min | ||||||||
Tune Type: | PFTBA | ||||||||
Ionization Mode: | EI | ||||||||
Sample Preparation | 100 µL of each standard was combined with 400 µL acetonitrile (final volume 1 mL) for a final concentration of 60 ppm. The sample was pipetted into a 2 mL short-cap, screw-thread vial (cat.# 21143) and capped with a short-cap, screw-vial closure (cat.# 24495). | ||||||||
Notes | Some compounds traditionally analyzed by LC were not GC amenable and/or only fragments were detected. These compounds, and the mixes that contain them, are: Abamectin (32586) (E)-Fenpyroximate (32586) Spinosad (32586) Aldicarb (32588) Hexythiazox (32588) Thiacloprid (32588) Clofentezine (32589) Daminozide (32589) |