Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.

Refinery Gas #6 on Rt-QS-BOND

GC_PC01014
Peaks
1.Methane
2.Ethane
3.Propylene
4.Propane
5.Isobutane
6.1-Butene
Peaks
7.Isobutylene
8.n-butane & cis-2-butene
9.trans-2-Butene
10.Isopentene
11.1-Pentene
12.n-pentane & cis-2-pentene
13.n-Hexane
ColumnRt-QS-BOND, 30 m, 0.53 mm ID, 20 µm (cat.# 19738)
Standard/Samplerefinery gas mix (mole%)
Injection
Inj. Vol.:20 µL split (split ratio 10:1)
Liner:Siltek Gooseneck Splitless (4mm)
Inj. Temp.:240 °C
Oven
Oven Temp.:40 °C (hold 2 min) to 225 °C at 15 °C/min (hold 5 min)
Carrier GasHe, constant flow
Flow Rate:5.7 mL/min
DetectorFID @ 240 °C
NotesSiltek Gooseneck Splitless (4mm) Obsolete, recommend catalog number 20709