Peaks | |
---|---|
1. | Methane |
2. | Ethane |
3. | Propane |
4. | Isobutane |
5. | n-Butane |
6. | trans-2-Butene |
7. | 1-Butene |
8. | Isobutylene |
Column | Rt-Alumina BOND / KCl, 50 m, 0.53 mm ID, 10 µm (cat.# 19760) |
---|---|
Standard/Sample | butane lighter fluid |
Injection | sample valve |
Sample Loop Vol.: | 100 µL |
Valve Name: | sample loop on gas sampling valve, split (split vent flow 400mL/min.), 2mm single gooseneck liner (cat.# 20795) |
Valve Temp.: | 200 °C |
Oven | |
Oven Temp.: | 45 °C (hold 1 min) to 200 °C at 10 °C/min (hold 3.5 min) |
Carrier Gas | H2, constant pressure (8.0 psi, 55.2 kPa) |
Linear Velocity: | 74 cm/sec @ 45 °C |
Detector | FID @ 200 °C |
---|