Peaks | |
---|---|
1. | Methanol |
2. | Dimethylamine |
3. | Trimethylamine |
4. | Methylethylamine |
5. | Dimethylethylamine |
6. | Diethylamine |
7. | Methyldiethylamine |
8. | Triethylamine |
Column | Rtx-Volatile Amine, 60 m, 0.32 mm ID (cat.# 18078) |
---|---|
Standard/Sample | Short-chain amines |
Diluent: | Water |
Injection | |
Inj. Vol.: | 1 µL split (split ratio 15:1) |
Inj. Temp.: | 220 °C |
Split Vent Flow Rate: | 30 mL/min |
Oven | |
Oven Temp.: | 40 °C (hold 10 min) to 250 °C at 20 °C/min (hold 10 min) |
Carrier Gas | H2, constant flow |
Flow Rate: | 2 mL/min |
Linear Velocity: | 35 cm/sec |
Detector | FID @ 250 °C |
---|---|
Instrument | Agilent/HP6890 GC |
Acknowledgement | Courtesy of Gilbert Baele, Taminco (Antwerp, Belgium) |