Peaks | |
---|---|
1. | Methanol |
2. | iso-Pentane |
3. | n-Pentane |
4. | trans-2-Pentene |
5. | tert-Butanol |
6. | cis-2-Pentene |
Column | Rtx-DHA-150, 150 m, 0.25 mm ID, 1.00 µm (cat.# 10149) |
---|---|
Standard/Sample | ASTM D5441 MTBE contaminants mix |
Conc.: | 1 % (w/w) each component in MTBE |
Injection | |
Inj. Vol.: | 0.2 µL split (split ratio 200:1) |
Liner: | Cup splitter (cat.# 20710) |
Inj. Temp.: | 200 °C |
Oven | |
Oven Temp.: | 60 °C (hold 13 min) to 180 °C at 10 °C/min (hold 20 min) |
Carrier Gas | He, constant flow |
Flow Rate: | 2.2 mL/min |
Linear Velocity: | 23 cm/sec @ 60 °C |
Detector | FID @ 250 °C |
---|---|
Make-up Gas Flow Rate: | 24 mL/min |
Make-up Gas Type: | H2 |
Hydrogen flow: | 30 mL/min |
Air flow: | 358 mL/min |
Instrument | Agilent 7890B GC |
Notes | Resolution between trans-2-pentene/tert-butanol and tert-butanol/cis-2-pentene meets the ASTM Method D5441 requirement of R ≥ 1.3. |