Simulated Distillation with Nitrogen Carrier Gas on MXT-1HT SimDist by ASTM D2887-B (Accelerated)

GC_PC1306
ColumnMXT-1HT SimDist, 10 m, 0.53 mm ID, 0.88 µm (cat.# 70134)
Standard/SampleASTM D2887-12 calibration standard (cat.# 31674)
Injection
Inj. Vol.:0.015 µL cool on-column
Temp. Program:80 °C to 360 °C at 35 °C/min
Oven
Oven Temp.:60 °C to 360 °C at 35 °C/min
Carrier GasN2, constant flow
Flow Rate:25.5 mL/min
DetectorFID @ 360 °C
Make-up Gas Flow Rate:20 mL/min
Make-up Gas Type:N2
Hydrogen flow:40 mL/min
Air flow:360 mL/min
InstrumentAgilent 7890B GC
NotesAccelerated analysis based on ASTM Method D2887 (Procedure B)