Peaks | tR (min) | Conc. (vol.%) | |
---|---|---|---|
1. | Methane | 6.88 | 8.50 |
2. | Ethene (ethylene) | 7.01 | 2.00 |
3. | Ethane | 7.09 | 6.00 |
4. | Propene (propylene) | 7.66 | 3.00 |
5. | Propane | 7.74 | 7.00 |
6. | 1,2-Propadiene | 8.18 | 0.85 |
7. | Isobutane | 8.97 | 6.00 |
Peaks | tR (min) | Conc. (vol.%) | |
---|---|---|---|
8. | Isobutene (isobutylene) | 9.77 | 2.00 |
9. | 1-Butene | 9.81 | 2.00 |
10. | 1,3-Butadiene | 9.94 | 3.00 |
11. | n-Butane | 10.17 | 4.00 |
12. | trans-2-Butene | 10.64 | 2.00 |
13. | cis-2-Butene | 11.37 | 2.00 |
14. | Isopentane | 15.13 | 1.00 |
15. | n-Pentane | 18.11 | 1.00 |
Column | Rtx-DHA-100, 100 m, 0.25 mm ID, 0.50 µm (cat.# 10148) |
---|---|
Standard/Sample | Refinery gas standard #1 (cat.# 34441) |
Injection | |
Inj. Vol.: | 30 µL split (split ratio 250:1) |
Liner: | Premium 4.0 mm ID Precision inlet liner w/wool (cat.# 23305.1) |
Inj. Temp.: | 250 °C |
Oven | |
Oven Temp.: | 0 °C (hold 15 min) to 50 °C at 1 °C/min |
Carrier Gas | He, constant flow |
Flow Rate: | 1.95 mL/min |
Linear Velocity: | 24.5 cm/sec @ 0 °C |
Detector | FID @ 300 °C |
---|---|
Make-up Gas Flow Rate: | 30 mL/min |
Make-up Gas Type: | N2 |
Hydrogen flow: | 40 mL/min |
Air flow: | 450 mL/min |
Instrument | Agilent 7890B GC |