Simulated Distillation of Gasoline on MXT-1 (ASTM D7096-16)

GC_PC1326
PeakstR (min)Conc.
(vol.%)
1.2-Methylbutane1.4679.6
2.n-Pentane1.6057.7
3.2-Methylpentane2.0805.8
4.n-Hexane2.3285.8
5.2,4-Dimethylpentane2.5845.8
6.n-Heptane3.1919.6
7.Toluene3.72111.5
8.n-Octane4.0454.8
PeakstR (min)Conc.
(vol.%)
9.p-Xylene4.58613.5
10.n-Propylbenzene5.2344.8
11.n-Decane5.5703.8
12.n-Butylbenzene5.9653.8
13.n-Dodecane6.8773.8
14.n-Tridecane7.4661.9
15.n-Tetradecane8.0171.9
16.n-Pentadecane8.5361.9
17.n-Hexadecane9.0253.8
ColumnMXT-1, 30 m, 0.53 mm ID, 5.00 µm (cat.# 70179)
Injection
Inj. Vol.:1 µL split (split ratio 50:1)
Liner:Topaz 4.0 mm ID Precision inlet liner w/wool (cat.# 23305)
Inj. Temp.:250 °C
Oven
Oven Temp.:40 °C (hold 1 min) to 245 °C at 25 °C/min (hold 4 min)
Carrier GasHe, flow program
Flow Program:5 mL/min (hold 0.5 min) to 20 mL/min at 60 mL/min/min
DetectorFID @ 250 °C
Constant Column + Constant Make-up:30 mL/min
Make-up Gas Type:He
Hydrogen flow:40 mL/min
Air flow:400 mL/min
InstrumentAgilent 7890B GC