Peaks | tR (min) | Conc. (vol.%) | |
---|---|---|---|
1. | Methanol | 1.089 | 4.1 |
2. | Ethanol | 1.302 | 4.1 |
3. | tert-Butanol | 1.670 | 4.1 |
4. | MTBE | 2.039 | 4.1 |
5. | n-Hexane | 2.329 | 6.1 |
6. | ETBE | 2.467 | 4.1 |
7. | 2,4-Dimethylpentane | 2.585 | 6.1 |
8. | TAME | 2.953 | 4.1 |
9. | n-Heptane | 3.192 | 10.2 |
10. | Toluene | 3.722 | 12.2 |
Peaks | tR (min) | Conc. (vol.%) | |
---|---|---|---|
11. | n-Octane | 4.046 | 5.1 |
12. | p-Xylene | 4.585 | 8.2 |
13. | n-Propylbenzene | 5.233 | 5.1 |
14. | n-Decane | 5.569 | 4.1 |
15. | n-Butylbenzene | 5.964 | 4.1 |
16. | n-Dodecane | 6.876 | 4.1 |
17. | n-Tridecane | 7.464 | 2.0 |
18. | n-Tetradecane | 8.015 | 2.0 |
19. | n-Pentadecane | 8.533 | 2.0 |
20. | n-Hexadecane | 9.023 | 4.1 |
Column | MXT-1, 30 m, 0.53 mm ID, 5.00 µm (cat.# 70179) |
---|---|
Standard/Sample | D7096 custom standard with oxygenates |
Injection | |
Inj. Vol.: | 1 µL split (split ratio 50:1) |
Liner: | Topaz 4.0 mm ID Precision inlet liner w/wool (cat.# 23305) |
Inj. Temp.: | 250 °C |
Oven | |
Oven Temp.: | 40 °C (hold 1 min) to 245 °C at 25 °C/min (hold 4 min) |
Carrier Gas | He, flow program |
Flow Program: | 5 mL/min (hold 0.5 min) to 20 mL/min at 60 mL/min/min |
Detector | FID @ 250 °C |
---|---|
Constant Column + Constant Make-up: | 30 mL/min |
Make-up Gas Type: | He |
Hydrogen flow: | 40 mL/min |
Air flow: | 400 mL/min |
Instrument | Agilent 7890B GC |