Peaks | tR (min) | Conc. (mg/mL) | Structural Nomenclature | |
---|---|---|---|---|
1. | Methyl capronate | 3.629 | 1.2 | C6:0 |
2. | Methyl caprylate | 6.237 | 1.2 | C8:0 |
3. | Methyl caprate | 8.787 | 1.2 | C10:0 |
4. | Methyl undecanoate | 9.971 | 0.6 | C11:0 |
5. | Methyl laurate | 11.105 | 1.2 | C12:0 |
6. | Methyl tridecanoate | 12.179 | 0.6 | C13:0 |
7. | Methyl myristate | 13.215 | 1.2 | C14:0 |
8. | Methyl myristoleate | 13.549 | 0.6 | C14:1 (cis-9) |
9. | Methyl pentadecanoate | 14.196 | 0.6 | C15:0 |
10. | Methyl pentadecenoate | 14.524 | 0.6 | C15:1 (cis-10) |
11. | Methyl palmitate | 15.152 | 1.8 | C16:0 |
12. | Methyl palmitoleate | 15.355 | 0.6 | C16:1 (cis-9) |
13. | Methyl margarate | 16.052 | 0.6 | C17:0 |
14. | Methyl heptadecenoate | 16.261 | 0.6 | C17:1 (cis-10) |
15. | Methyl stearate | 16.995 | 1.2 | C18:0 |
16. | Methyl oleate | 17.156 | 1.2 | C18:1 (cis-9) |
17. | Methyl elaidate | 17.168 | 1.2 | C18:1 (trans-9) |
18. | Methyl linoleate | 17.583 | 0.6 | C18:2 (all-cis-9,12) |
Peaks | tR (min) | Conc. (mg/mL) | Structural Nomenclature | |
---|---|---|---|---|
19. | Methyl linolelaidate | 17.641 | 0.6 | C18:2 (all-trans-9,12) |
20. | Methyl γ-linolenate | 17.874 | 0.6 | C18:3 (all-cis-6,9,12) |
21. | Methyl nonadecanoate | 18.052 | 2.0 | C19:0 |
22. | Methyl α-linolenate | 18.223 | 0.6 | C18:3 (all-cis-9,12,15) |
23. | Methyl arachidate | 19.075 | 1.2 | C20:0 |
24. | Methyl (Z)-11-eicosenoate | 19.255 | 0.6 | C20:1 (cis-11) |
25. | Methyl 11,14-eicosadienoate | 19.761 | 0.6 | C20:2 (all-cis-11,14) |
26. | Methyl eicosa-8,11,14-trienoate | 20.046 | 0.6 | C20:3 (all-cis-8,11,14) |
27. | Methyl heneicosanoate | 20.197 | 0.6 | C21:0 |
28. | Methyl arachidonate | 20.290 | 0.6 | C20:4 (all-cis-5,8,11,14) |
29. | Methyl 11,14,17-eicosatrienoate | 20.488 | 0.6 | C20:3 (all-cis-11,14,17) |
30. | Methyl 5,8,11,14,17-eicosapentanoate | 21.036 | 0.6 | C20:5 (all-cis-5,8,11,14,17) |
31. | Methyl behenate | 21.39 | 1.2 | C22:0 |
32. | Methyl erucate | 21.595 | 0.6 | C22:1 (cis-13) |
33. | Methyl docosadienoate | 22.150 | 0.6 | C22:2 (all-cis-13,16) |
34. | Methyl tricosanoate | 22.584 | 0.6 | C23:0 |
35. | Methyl lignocerate | 23.826 | 1.2 | C24:0 |
36. | Methyl docosahexaenoate | 23.863 | 0.6 | C22:6 (all-cis-4,7,10,13,16,19) |
37. | Methyl nervonate | 24.055 | 0.6 | C24:1 (cis-15) |
Column | FAMEWAX, 30 m, 0.25 mm ID, 0.25 µm (cat.# 12497) |
---|---|
Standard/Sample | Food industry FAME mix (cat.# 35077) |
Methyl nonadecanoate (cat.# 35055) | |
Diluent: | Standard cat.# 35055 was dissolved in toluene. |
Injection | |
Inj. Vol.: | 1 µL split (split ratio 100:1) |
Liner: | Topaz 4.0 mm ID Precision inlet liner w/wool (cat.# 23305) |
Inj. Temp.: | 240 °C |
Oven | |
Oven Temp.: | 60 °C (hold 2 min) to 200 °C at 10 °C/min to 240 °C at 5 °C/min (hold 7 min) |
Carrier Gas | H2, constant flow |
Flow Rate: | 1.7 mL/min |
Detector | FID @ 250 °C |
---|---|
Instrument | Agilent 7890B GC |
Notes | This chromatogram is an overlay of two injections: food industry FAME standard (black) and C19:0 methyl ester in toluene (red). An excellent separation of C19:0 (used in EN 14103 as an internal standard) and the most prevalent FAMEs found in biodiesel blends was achieved. Note that C4:0 from the food industry FAME standard elutes in the solvent front. |