Peaks | Mol% | |
---|---|---|
1. | Nitrogen | Balance |
2. | Carbon monoxide | 1.00 |
3. | Methane | 5.00 |
4. | Carbon dioxide | 3.00 |
5. | Ethylene | 2.00 |
6. | Acetylene | 1.00 |
7. | Ethane | 4.00 |
8. | Propylene | 3.00 |
9. | Propane | 6.00 |
10. | Propadiene | 1.00 |
11. | Cyclopropane | 0.04 |
12. | Isobutane | 5.00 |
Peaks | Mol% | |
---|---|---|
13. | Isobutene | 1.00 |
14. | Butene-1 | 2.00 |
15. | 1,3-Butadiene | 3.00 |
16. | Butane | 4.00 |
17. | trans-2-Butene | 3.00 |
18. | cis-2-Butene | 2.00 |
19. | Isopentane | 1.00 |
20. | Pentene-1 | 0.40 |
21. | trans-2-Pentene | 0.40 |
22. | 2-Methyl-2-butene | 0.20 |
23. | Pentane | 1.00 |
24. | cis-2-Pentene | 0.15 |
25. | Hexane | 0.10 |
Column | Rt-Q-BOND, 30 m, 0.53 mm ID, 20 µm (cat.# 19742) |
---|---|
Standard/Sample | Refinery gas standard #2 (cat.# 34442) |
Injection | split |
Liner: | Topaz 2.0 mm ID straight inlet liner (cat.# 23313) |
Inj. Temp.: | 200 °C |
Oven | |
Oven Temp.: | 40 °C to 250 °C at 10 °C/min |
Carrier Gas | He, constant flow |
Flow Rate: | 4 mL/min |
Detector | FID |
---|---|
Instrument | Agilent 7890B GC |
Notes | The sample was analyzed using two detectors: 1. FID at 260 °C, 40 µL split injection (split ratio 10:1) 2. TCD at 250°C, 150 µL split injection (split ratio 10:1) |