Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.
Peaks | |
---|---|
1. | Benzene |
2. | Toluene |
3. | Ethylbenzene |
4. | p-Xylene |
5. | m-Xylene |
6. | Cumene |
7. | o-Xylene |
Peaks | |
---|---|
8. | Propylbenzene |
9. | p-Ethyltoluene |
10. | m-Ethyltoluene |
11. | Styrene |
12. | Allylbenzene |
13. | α-Methylstyrene |
14. | Phenylacetylene |
15. | trans-β-Methylstyrene |
Column | Stabilwax, 60 m, 0.32 mm ID, 0.25 µm (cat.# 10627) |
---|---|
Standard/Sample | Styrene |
Injection | |
Inj. Vol.: | 0.6 µL split (split ratio 100:1) |
Liner: | Topaz 4.0 mm ID Precision inlet liner w/wool (cat.# 23305) |
Inj. Temp.: | 250 °C |
Oven | |
Oven Temp.: | 60 °C (hold 10 min) to 150 °C at 5 °C/min (hold 4 min) |
Carrier Gas | He, constant flow |
Flow Rate: | 1.7 mL/min |
Detector | FID @ 265 °C |
---|---|
Instrument | Agilent 7890B GC |
Notes | Purity analysis of styrene. |
We have determined your location as:
Or select your location below:
Sorry! We are unable to determine your location. Please select your location below.