Peaks | tR (min) | |
---|---|---|
1. | Octane | 4.79 |
2. | Nonane | 5.60 |
3. | Benzene | 6.32 |
4. | Decane | 7.19 |
5. | Toluene | 8.55 |
6. | 1,4-Dioxane | 9.24 |
7. | Undecane | 10.48 |
8. | Ethylbenzene | 11.90 |
9. | p-Xylene | 12.25 |
10. | m-Xylene | 12.52 |
Peaks | tR (min) | |
---|---|---|
11. | Cumene | 13.88 |
12. | o-Xylene | 14.37 |
13. | Dodecane | 14.81 |
14. | Propylbenzene | 15.35 |
15. | p-Ethyltoluene | 15.92 |
16. | m-Ethyltoluene | 16.02 |
17. | tert-Butylbenzene | 16.49 |
18. | Isobutylbenzene | 16.62 |
19. | Mesitylene | 16.77 |
20. | sec-Butylbenzene | 16.94 |
Peaks | tR (min) | |
---|---|---|
21. | Styrene | 17.32 |
22. | Allylbenzene | 17.60 |
23. | 1,2,4-Trimethylbenzene | 18.22 |
24. | Tridecane | 18.75 |
25. | 1,3-Diethylbenzene | 18.93 |
26. | 1,4-Diethylbenzene | 19.25 |
27. | n-Butylbenzene | 19.35 |
28. | α-Methylstyrene | 20.04 |
29. | Phenylacetylene | 21.12 |
30. | 1,4-Diisopropylbenzene | 22.25 |
31. | trans-β-Methylstyrene | 22.46 |
Column | Stabilwax, 60 m, 0.32 mm ID, 0.25 µm (cat.# 10627) |
---|---|
Standard/Sample | D7504 resolution mix |
Diluent: | Cyclohexane |
Injection | |
Inj. Vol.: | 0.6 µL split (split ratio 100:1) |
Liner: | Topaz 4.0 mm ID Precision inlet liner w/wool (cat.# 23305) |
Inj. Temp.: | 250 °C |
Oven | |
Oven Temp.: | 60 °C (hold 10 min) to 150 °C at 5 °C/min (hold 4 min) |
Carrier Gas | He, constant flow |
Flow Rate: | 1.7 mL/min |
Detector | FID @ 265 °C |
---|---|
Instrument | Agilent 7890B GC |
Notes | ASTM D7504 covers the determination of trace impurities in monocyclic aromatic hydrocarbons. It includes analysis of nonaromatic and individual aromatic compounds in benzene, toluene, ethylbenzene, p-xylene, o-xylene, styrene, and mixed xylenes. The chromatogram shows a synthetic blend of the most common analytes identified and quantified by D7504; all analytes are well resolved on a Stabilwax column. |