Peaks | |
---|---|
1. | Ethanol |
2. | 2-Methylpentane |
3. | Isooctane |
4. | Toluene |
5. | Octane |
Peaks | |
---|---|
6. | Ethylbenzene |
7. | m-Xylene |
8. | p-Xylene |
9. | o-Xylene |
10. | 1,2,4-Trimethylbenzene |
11. | Methylnaphthalenes |
Column | Rxi-1ms, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13323) |
---|---|
Standard/Sample | Gasoline |
Conc.: | Neat |
Injection | |
Inj. Vol.: | 1 µL split (split ratio 300:1) |
Liner: | Topaz 4.0 mm ID Precision inlet liner w/wool (cat.# 23305) |
Inj. Temp.: | 250 °C |
Oven | |
Oven Temp.: | 35 °C (hold 10 min) to 200 °C at 7 °C/min |
Carrier Gas | He, constant flow |
Flow Rate: | 1 mL/min |
Detector | VUV |
---|---|
Transfer Line Temp.: | 275 °C |
Flow Cell Temp.: | 275 °C |
Acquisition Range: | 125-240 nm |
Acquisition Rate: | 4.5 spectra/sec |
Instrument | HP 6890 GC & VUV Analytics VGA-100 |
Notes | ASTM D8071: Standard Test Method for Determination of Hydrocarbon Group Types and Select Hydrocarbon and Oxygenate Compounds in Automotive Spark-Ignition Engine Fuel Using Gas Chromatography with Vacuum Ultraviolet Absorption Spectroscopy Detection (GC-VUV) |
Acknowledgement | Restek would like to thank Jack Cochran from VUV Analytics for sharing the data. |