Peaks | |
---|---|
1. | Helium |
2. | Neon |
3. | Hydrogen |
4. | Oxygen |
5. | Argon |
6. | Nitrogen |
Peaks | |
---|---|
7. | Carbon monoxide |
8. | Methane |
9. | Krypton |
10. | Carbon dioxide |
11. | Xenon |
12. | Sulfur hexafluoride |
Column | 100/120 mesh, ShinCarbon ST micropacked column, 2 m, 1.00 mm ID (cat.# 19808) |
---|---|
Standard/Sample | Noble and permanent gas standard |
Conc.: | 2-10 mol% |
Injection | |
Inj. Vol.: | 200 µL split (split ratio 10:1) |
Liner: | Topaz 2.0 mm ID straight inlet liner (cat.# 23313) |
Inj. Temp.: | 150 °C |
Oven | |
Oven Temp.: | 35 °C (hold 3 min) to 80 °C at 10 °C/min to 200 °C at 30 °C/min (hold 5 min) |
Carrier Gas | He, constant flow |
Flow Rate: | 6.5 mL/min |
Detector | TCD @ 250 °C |
---|---|
Instrument | Agilent 7890B GC |
Notes | Chromatogram A Analysis Conditions: Inj. Vol.: 50 µL, split (split ratio 10:1) Conc.: 33 mol% each Inj. Temp.: 150 °C Oven Temp.: 35 °C isothermal Carrier Gas: N2, constant flow Flow Rate: 5.5 mL/min Detector: TCD @ 250 °C |