Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.
Peaks | tR (min) | |
---|---|---|
1. | Glycerol | 4.682 |
2. | 1,2,4-Butanetriol | 5.624 |
3. | Monoglycerides | as shown |
4. | Tricaprin | 19.174 |
5. | Diglycerides | 20.13 - 20.80 |
6. | Triglycerides | 22.24 - 24.70 |
Column | Rtx-Biodiesel TG with 2 m x 0.53 mm ID retention gap, 10 m, 0.32 mm ID, 0.1 µm (cat.# 10291) |
---|---|
Standard/Sample | Soy B100 biodiesel |
Tricaprin (1,2,3-tricaprinoylglycerol) (cat.# 33025) | |
(s)-(-)-1,2,4-Butanetriol (cat.# 33024) | |
Diluent: | Heptane |
Injection | |
Inj. Vol.: | 1 µL cool on-column |
Temp. Program: | Track oven |
Oven | |
Oven Temp.: | 50 °C (hold 1 min) to 180 °C at 15 °C/min to 230 °C at 7 °C/min to 380 °C at 30 °C/min (hold 10 min) |
Carrier Gas | H2, constant flow |
Flow Rate: | 1.6 mL/min |
Detector | FID @ 380 °C |
---|---|
Instrument | Agilent 7890B GC |
Sample Preparation | Sample preparation followed ASTM D6584-17: Soy biodiesel B100 was fortified with internal standards (butanetriol and tricaprin) and derivatized using MSTFA. |
We have determined your location as:
Or select your location below:
Sorry! We are unable to determine your location. Please select your location below.