Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.
Peaks | tR (min) | |
---|---|---|
1. | 3-Methyl-1-pentene | 14.06 |
2. | 3-Methylpentane | 16.31 |
3. | 1-Hexene | 18.04 |
4. | Hexane | 18.39 |
5. | 2-Ethyl-1-butene | 19.08 |
6. | cis-3-Hexene | 19.61 |
Peaks | tR (min) | |
---|---|---|
7. | trans-2-Hexene | 19.61 |
8. | cis-3-Methyl-2-pentene | 20.41 |
9. | cis-2-Hexene | 20.81 |
10. | trans-3-Methyl-2-pentene | 21.19 |
11. | Methyl-cyclopentene | 22.87 |
12. | Cyclohexene | 23.24 |
Column | Rxi-LAO, 60 m, 0.25 mm ID, 1.4 µm (cat.# 13876) |
---|---|
Standard/Sample | 1-Hexene |
Conc.: | Neat solvent |
Injection | |
Inj. Vol.: | 1 µL split (split ratio 100:1) |
Liner: | Topaz 4.0 mm ID low pressure drop Precision inlet liner w/wool (cat.# 23309) |
Inj. Temp.: | 250 °C |
Split Vent Flow Rate: | 125 mL/min |
Oven | |
Oven Temp.: | 35 °C (hold 20 min) to 160 °C at 30 °C/min (hold 20 min) |
Carrier Gas | He, constant flow |
Linear Velocity: | 23 cm/sec @ 35 °C |
Detector | FID @ 300 °C |
---|---|
Make-up Gas Flow Rate: | 45 mL/min |
Make-up Gas Type: | N2 |
Hydrogen flow: | 40 mL/min |
Air flow: | 450 mL/min |
Data Rate: | 20 Hz |
Instrument | Agilent 7890B GC |
Sample Preparation | The sample was pipetted into a 2 mL vial (cat.# 21142) and capped with a short screw cap (cat.# 24498). |
Notes | • Compounds were tentatively identified using a mass spectrometer and method translation. • A 208V instrument was used. • Benzene is not present in the sample, but if it were, it would elute at 23.73 minutes under these conditions. |
We have determined your location as:
Or select your location below:
Sorry! We are unable to determine your location. Please select your location below.