1-Hexene on Rxi-LAO (40 m x 0.18 mm x 1.0 µm)

GC_PC1380
PeakstR (min)
1.3-Methyl-1-pentene5.25
2.3-Methylpentane6.10
3.1-Hexene6.87
4.Hexane6.94
5.2-Ethyl-1-butene7.16
6.cis-3-Hexene7.34
7.trans-2-Hexene7.35
8.cis-3-Methyl-2-pentene7.66
9.cis-2-Hexene7.91
10.trans-3-Methyl-2-pentene8.23
ColumnRxi-LAO, 40 m, 0.18 mm ID, 1.0 µm (cat.# 40815)
Standard/Sample1-Hexene
Conc.: Neat solvent
Injection
Inj. Vol.:1 µL split (split ratio 100:1)
Liner:Topaz 4.0 mm ID low pressure drop Precision inlet liner w/wool (cat.# 23309)
Inj. Temp.:250 °C
Split Vent Flow Rate:178 mL/min
Oven
Oven Temp.:35 °C (hold 11.85 min) to 160 °C at 50 °C/min (hold 12 min)
Carrier GasHe, constant flow
Linear Velocity:39.19 cm/sec @ 35 °C
DetectorFID @ 300 °C
Make-up Gas Flow Rate:45 mL/min
Make-up Gas Type:N2
Hydrogen flow:40 mL/min
Air flow:400 mL/min
Data Rate:20 Hz
InstrumentAgilent 7890B GC
Sample PreparationThe sample was pipetted into a 2 mL vial (cat.# 21142) and capped with a short screw cap (cat.# 24498).
Notes• Compounds were tentatively identified using a mass spectrometer and method translation.
• A 208V instrument was used.
• Benzene is not present in the sample, but if it were, it would elute at 12.36 minutes under these conditions.