Column | Rxi-LAO, 40 m, 0.18 mm ID, 1.0 µm (cat.# 40815) |
---|---|
Standard/Sample | 1-Octene |
Conc.: | Neat solvent |
Injection | |
Inj. Vol.: | 1 µL split (split ratio 100:1) |
Liner: | Topaz 4.0 mm ID low pressure drop Precision inlet liner w/wool (cat.# 23309) |
Inj. Temp.: | 250 °C |
Split Vent Flow Rate: | 178 mL/min |
Oven | |
Oven Temp.: | 35 °C (hold 11.85 min) to 160 °C at 50 °C/min (hold 12 min) |
Carrier Gas | He, constant flow |
Linear Velocity: | 39.19 cm/sec @ 35 °C |
Detector | FID @ 300 °C |
---|---|
Make-up Gas Flow Rate: | 45 mL/min |
Make-up Gas Type: | N2 |
Hydrogen flow: | 40 mL/min |
Air flow: | 400 mL/min |
Data Rate: | 20 Hz |
Instrument | Agilent 7890B GC |
Sample Preparation | The sample was pipetted into a 2 mL vial (cat.# 21142) and capped with a short screw cap (cat.# 24498). |
Notes | • Peaks are tentatively identified. • A 208V instrument was used. |