Peaks | |
---|---|
1. | Methanol |
2. | Acetaldehyde |
3. | Ethanol |
4. | Isopropanol |
5. | Acetone |
6. | n-Propanol |
Columns | Rtx-BAC1 30 m, 0.53 mm ID, 3.00 µm (cat.# 18001) |
---|---|
and Rtx-BAC2 30 m, 0.53 mm ID, 2.00 µm (cat.# 18000) | |
Standard/Sample | 1.0 mL headspace sample of a blood alcohol mix |
Conc.: | 0.1% per compound |
Injection | |
Inj. Vol.: | 1,000 µL headspace-syringe split |
Inj. Temp.: | 200 °C |
Oven | |
Oven Temp.: | 40 °C |
Carrier Gas | He, constant pressure |
Linear Velocity: | 80 cm/sec @ 40 °C |
Detector | FID @ 200 °C |
---|---|
Notes | Baseline resolution in less than 3 minutes! |