Peaks | |
---|---|
1. | Ethosuximide |
2. | Barbital |
3. | Methyprylon |
4. | Aprobarbital |
5. | Butalbital |
6. | Amobarbital |
7. | Pentobarbital |
8. | Secobarbital |
Peaks | |
---|---|
9. | Meprobamate |
10. | Carisoprodal |
11. | Glutethimide |
12. | Phenobarbital |
13. | Methaqualone |
14. | Primidone |
15. | Carbamazepine |
16. | Diphenylhydantoin |
Column | Rtx-35, 30 m, 0.53 mm ID, 1.00 µm (cat.# 10455) |
---|---|
Standard/Sample | Acidic/neutral drugs |
Conc.: | 50 µg/mL |
Injection | |
Inj. Vol.: | 1.0 µL splitless (hold 0.5 min) |
Inj. Temp.: | 250 °C |
Oven | |
Oven Temp.: | 100 °C to 280 °C at 10 °C/min (hold 5 min) |
Carrier Gas | He, constant pressure |
Linear Velocity: | 40 cm/sec @ 100 °C |
Detector | FID @ 250 °C |
---|---|
Notes | FID sensitivity: 5.12 x 10-10 AFS |