Peaks | |
---|---|
12. | Pseudoephedrine - HFBA |
13. | Phenmetrazine - HFBA |
14. | Phendimetrazine - HFBA |
15. | Methylenedioxyamphetamine - HFBA |
16. | Diethylpropion - HFBA |
17. | Methylene dioxymethamphetamine - HFBA |
18. | Methylene dioxyethylamphetamine - HFBA |
19. | 4-Methyl-2,5-dimethoxyamphetamine - HFBA |
20. | Phenylephrine - HFBA |
21. | Caffeine - HFBA |
22. | Benzphetamine - HFBA |
Column | Rtx-200, 30 m, 0.25 mm ID, 0.50 µm (cat.# 15038) |
---|---|
Injection | |
Inj. Vol.: | 1 µL split |
Inj. Temp.: | 250 °C |
Split Vent Flow Rate: | 45 mL/min |
Oven | |
Oven Temp.: | 100 °C to 310 °C at 10 °C/min |
Carrier Gas | He |
Detector | MS |
---|---|
Mode: | Scan |
Ionization Mode: | EI |
Scan Range: | 40-600 amu |
Instrument | Varian 3400 coupled with with Varian Saturn 2000 MS detector |