Column | Stabilwax, 30 m, 0.32 mm ID, 0.25 µm (cat.# 10624) |
---|---|
Standard/Sample | Residual solvents - class 1 (cat.# 36279) |
in 20 mL headspace vial (cat.# 24685) | |
Diluent: | Water |
Injection | |
Inj. Vol.: | 1.0 µL headspace-syringe split (split ratio 5:1) |
Liner: | 2 mm splitless (cat.# 20712) |
Inj. Temp.: | 140 °C |
Headspace-Syringe | |
Instrument: | Overbrook Scientific HT200H |
Syringe Temp.: | 100 °C |
Sample Temp.: | 80 °C |
Sample Equil. Time: | 45 min |
Inj. Speed: | 8 |
Inj. Dwell: | 5 sec. |
Oven | |
Oven Temp.: | 50 °C (hold 20 min) to 165 °C at 6 °C/min (hold 20 min) |
Carrier Gas | He, constant flow |
Flow Rate: | 2.15 mL/min |
Linear Velocity: | 35.2 cm/sec |
Detector | FID @ 250 °C |
---|