Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.
Peaks | |
---|---|
1. | Cyclohexane |
2. | Methylcyclohexane |
3. | trans-1,2-Dichloroethene |
4. | Tetrahydrofuran |
5. | Methanol |
6. | Dichloromethane |
7. | cis-1,2-Dichloroethene |
Peaks | |
---|---|
8. | Acetonitrile |
9. | Toluene |
10. | 1,4-Dioxane |
11. | Ethyl benzene |
12. | p-Xylene |
13. | m-Xylene |
14. | o-Xylene |
15. | Chlorobenzene |
Column | Stabilwax, 30 m, 0.32 mm ID, 0.25 µm (cat.# 10624) |
---|---|
Standard/Sample | Residual solvents class 2 - mix A (cat.# 36271) |
in 20 mL headspace vial (cat.# 24685) | |
Diluent: | Water |
Injection | |
Inj. Vol.: | 1,000 µL headspace-syringe split (split ratio 5:1) |
Liner: | 2 mm splitless (cat.# 20712) |
Inj. Temp.: | 140 °C |
Headspace-Syringe | |
Instrument: | Overbrook Scientific HT200H |
Syringe Temp.: | 100 °C |
Sample Temp.: | 80 °C |
Sample Equil. Time: | 45 min |
Inj. Speed: | 8 |
Inj. Dwell: | 5 sec. |
Oven | |
Oven Temp.: | 50 °C (hold 20 min) to 165 °C at 6 °C/min (hold 20 min) |
Carrier Gas | He, constant flow |
Flow Rate: | 2.15 mL/min |
Linear Velocity: | 35.2 cm/sec |
Detector | FID @ 250 °C |
---|
We have determined your location as:
Or select your location below:
Sorry! We are unable to determine your location. Please select your location below.