Peaks | |
---|---|
1. | Amphetamine |
2. | Methamphetamine |
3. | MDA |
4. | MDMA |
5. | MDEA |
Column | Rtx®-5MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 12623) |
---|---|
Standard/Sample | amphetamine and methamphetamine |
Diluent: | methanol |
Conc.: | 100 µg/mL |
Injection | |
Inj. Vol.: | 1.0 µL split (split ratio 10:1) |
Liner: | Gooseneck Splitless (4mm) w/Wool Siltek Liner |
Inj. Temp.: | 250 °C |
Oven | |
Oven Temp.: | 70 °C to 250 °C at 15 °C/min (hold 5 min) |
Carrier Gas | H2, constant flow |
Flow Rate: | 1 mL/min |
Detector | FID @ 300 °C |
---|---|
Notes | Gooseneck Splitless (4mm) w/Wool Siltek Liner obsolete, recommend catalog number 20798 |