Peaks | tR (min) | Ions Monitored (m/z) | |
---|---|---|---|
1. | Methyl methanesulfonate (MMS) | 5.043 | 79, 80, 95 |
2. | Isopropyl methanesulfonate (iPMS) | 6.010 | 59, 79, 123 |
3. | Diethyl sulfate (DES) | 6.357 | 111, 125, 139 |
4. | di-isopropyl sulfate (DPS) | 6.970 | 45, 87, 167 |
5. | dibutyl sulfate (DBS) | 9.260 | 29, 41, 56 |
6. | Methyl benzenesulfonate (MBS) | 9.463 | 77, 141, 172 |
Peaks | tR (min) | Ions Monitored (m/z) | |
---|---|---|---|
7. | Ethyl benzenesulfonate (EBS) | 9.883 | 77, 141, 186 |
8. | Methyl toluenesulfonate (MTS) | 10.337 | 91, 155, 186 |
9. | Ethyl toluenesulfonate (ETS) | 10.713 | 91, 155, 200 |
10. | n-propyl toluenesulfonate (nPTS) | 10.827 | 91, 155, 172 |
11. | n-butyl benzenesulfonate (nBBS) | 11.07 | 77, 141, 159 |
12. | isopropyl p-toluenesulfonate (iPTS) | 11.240 | 91, 155, 172 |
13. | p-toluenesulfonic acid n-butyl ester (nBTS) | 11.823 | 91, 155, 173 |
Column | Rxi-624Sil MS, 20 m, 0.18 mm ID, 1.00 µm (cat.# 13865) |
---|---|
Standard/Sample | |
Diluent: | 90:10 acetonitrile:water |
Conc.: | 500 ng/mL |
Injection | |
Inj. Vol.: | 1 µL splitless (hold 0.5 min) |
Liner: | 3.5mm Single Gooseneck Liner with wool placed 3cm from top (middle) (cat.# 22286) |
Inj. Temp.: | 220 °C |
Purge Flow: | 3 mL/min |
Oven | |
Oven Temp.: | 80 °C (hold 2 min) to 300 °C at 20 °C/min (hold 5 min) |
Carrier Gas | He, constant flow |
Linear Velocity: | 45 cm/sec |
Detector | MS |
---|---|
Mode: | SIM |
Transfer Line Temp.: | 280 °C |
Analyzer Type: | Quadrupole |
Source Temp.: | 280 °C |
Solvent Delay Time: | 4 min |
Ionization Mode: | EI |
Instrument | Shimadzu 2010 GC & QP2010+ MS |
Acknowledgement | In collaboration with Merck and Company |