Peaks | tR (min) | |
---|---|---|
1. | Cyclohexane | 1.70 |
2. | Methylcyclohexane | 1.79 |
3. | trans-1,2-Dichloroethene | 2.08 |
4. | Tetrahydrofuran | 2.08 |
5. | Methanol | 2.25 |
6. | Dichloromethane | 2.52 |
7. | cis-1,2-Dichloroethene | 3.12 |
8. | Acetonitrile | 3.31 |
Peaks | tR (min) | |
---|---|---|
9. | Toluene | 3.82 |
10. | 1,4-Dioxane | 4.21 |
11. | Ethylbenzene | 5.76 |
12. | p-Xylene | 6.02 |
13. | m-Xylene | 6.24 |
14. | Cumene | 7.53 |
15. | o-Xylene | 8.00 |
16. | Chlorobenzene | 9.59 |
Column | Stabilwax, 30 m, 0.32 mm ID, 0.25 µm (cat.# 10624) |
---|---|
Standard/Sample | Residual Solvents Class 2 - Mix A (2013 Rev) (cat.# 36012) |
Diluent: | Water |
Conc.: | Per USP <467> method |
Injection | |
Inj. Vol.: | 1,000 µL headspace-loop split (split ratio 50:1) |
Liner: | Premium 1.0 mm ID straight inlet liner (cat.# 23333.1) |
Headspace-Loop | |
Inj. Port Temp.: | 140 °C |
Instrument: | Tekmar HT3 |
Inj. Time: | 1.0 min |
Transfer Line Temp.: | 105 °C |
Valve Oven Temp.: | 105 °C |
Needle Temp.: | 80 °C |
Standby flow rate: | 50 mL/min |
Sample Temp.: | 80 °C |
Platen temp equil. time: | 1.0 min |
Sample Equil. Time: | 45 min |
Mixer time: | 5.0 min |
Mixing level: | 5 |
Mixer stabilize time: | 0.50 min |
Vial Pressure: | 20 psi |
Pressurize Time: | 5.0 min |
Pressure Equilibration Time: | 0.20 min |
Loop Pressure: | 10 psi |
Loop Fill Time: | 2.0 min |
Oven | |
Oven Temp.: | 50 °C (hold 20 min) to 165 °C at 6 °C/min (hold 20 min) |
Carrier Gas | He, constant flow |
Flow Rate: | 2.2 mL/min |
Linear Velocity: | 36 cm/sec @ 50 °C |
Detector | FID @ 250 °C |
---|---|
Make-up Gas Flow Rate: | 45 mL/min |
Make-up Gas Type: | N2 |
Hydrogen flow: | 40 mL/min |
Air flow: | 450 mL/min |
Data Rate: | 20 Hz |
Instrument | Agilent/HP6890 GC |