Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.
Peaks | |
---|---|
1. | Cephradine |
Column | Ultra IBD (cat.# 9175552) |
---|---|
Dimensions: | 50 mm x 2.1 mm ID |
Particle Size: | 5 µm |
Pore Size: | 100 Å |
Temp.: | Ambient |
Standard/Sample | |
Diluent: | Mobile phase |
Conc.: | 5 µg/mL |
Inj. Vol.: | 5 µL |
Mobile Phase | 5 mM Ammonium acetate, pH 7.4:acetonitrile (20:80) |
Flow: | 0.2 mL/min |
Detector | PE/Sciex API 150 EX |
---|---|
Ion Source: | TurboIonSpray® |
Ion Mode: | ESI+ |
Ion Spray Voltage: | 4.7 kV |
Orifice Voltage: | 10.0 V |
Ring Voltage: | 25.0 V |
Source Temp.: | 350 °C |
Notes | Nebulizer: 8L/hour Curtain gas: 12L/hour SIM monitoring 349 ± 3 Daltons |
We have determined your location as:
Or select your location below:
Sorry! We are unable to determine your location. Please select your location below.