Peaks | |
---|---|
1. | Penicillin V |
Column | Ultra IBD (cat.# 9175552) |
---|---|
Dimensions: | 50 mm x 2.1 mm ID |
Particle Size: | 5 µm |
Pore Size: | 100 Å |
Temp.: | Ambient |
Standard/Sample | |
Diluent: | Mobile phase |
Conc.: | 5 µg/mL |
Inj. Vol.: | 5 µL |
Mobile Phase | 5 mM Ammonium acetate pH 7.4:acetonitrile (20:80) |
Flow: | 0.2 mL/min |
Detector | PE/Sciex API 150 EX |
---|---|
Ion Source: | TurboIonSpray® |
Ion Mode: | ESI+ |
Ion Spray Voltage: | 4.7 kV |
Orifice Voltage: | 10.0 V |
Ring Voltage: | 25.0 V |
Source Temp.: | 350 °C |
Notes | Nebulizer: 8L/hour Curtain gas: 12L/hour SIM monitoring: 349 ± 3dal. |