Peaks | |
---|---|
1. | Aldicarb sulfone |
2. | Aldicarb sulfoxide |
3. | Oxamyl |
4. | Methomyl |
5. | 3-Hydroxycarbofuran |
Peaks | |
---|---|
6. | Aldicarb |
7. | Propoxur |
8. | Carbofuran |
9. | Carbaryl |
10. | Methiocarb |
11. | 4-Bromo-3,5-dimethylcarbamate |
Column | Ultra Carbamate (cat.# 9177355) | ||||||
---|---|---|---|---|---|---|---|
Dimensions: | 50 mm x 4.6 mm ID | ||||||
Particle Size: | 3 µm | ||||||
Pore Size: | 100 Å | ||||||
Temp.: | 27 °C | ||||||
Standard/Sample | 4-Bromo-3,5-dimethylphenyl-N-methylcarbamate (BDMC) (cat.# 32274) | ||||||
531.1 Carbamate pesticide calibration mixture (cat.# 32273) | |||||||
Diluent: | Methanol | ||||||
Conc.: | 50 µg/mL each | ||||||
Inj. Vol.: | 5 µL | ||||||
Mobile Phase | |||||||
A: | 90:10 Water:methanol | ||||||
B: | 90:10 Methanol:acetonitrile | ||||||
| |||||||
Flow: | 1.5 mL/min |
Detector | UV/Vis @ 220 nm |
---|---|
Notes | Standards mixed 50:50 |