Peaks | tR (min) | MRM1 | MRM2 | MRM3 | |
---|---|---|---|---|---|
1. | 7-Aminoclonazepam | 0.57 | 286.0/250.3 | 286.0/222.2 | 286.0/121.0 |
2. | 7-Aminoflunitrazepam | 0.74 | 284.2/135.2 | 284.2/240.2 | 284.2/226.1 |
3. | Bromazepam | 1.07 | 315.9/182.0 | 315.9/209.1 | 315.9/260.9 |
4. | N-Desmethylflunitrazepam | 1.70 | 300.0/254.1 | 300.0/225.2 | 300.0/198.0 |
5. | Nitrazepam | 1.68 | 282.1/236.1 | 282.1/207.2 | 282.1/180.0 |
6. | Lorazepam | 1.70 | 321.0/302.9 | 321.0/275.1 | 321.0/229.1 |
7. | 2-Hydroxyethylflurazepam | 1.78 | 333.1/246.1 | 333.1/166.0 | 333.1/109.1 |
8. | Nordiazepam | 1.76 | 271.0/140.0 | 271.0/208.0 | 271.0/225.9 |
9. | Estazolam | 1.86 | 294.9/267.1 | 294.9/241.0 | 294.9/138.0 |
10. | Desalkylflurazepam | 1.90 | 289.0/140.0 | 289.0/165.1 | 289.0/226.0 |
Peaks | tR (min) | MRM1 | MRM2 | MRM3 | |
---|---|---|---|---|---|
11. | Clonazepam | 1.91 | 315.9/270.2 | 315.9/214.1 | 315.9/207.1 |
12. | Alprazolam | 1.98 | 309.4/281.1 | 309.4/205.0 | 309.4/274.1 |
13. | Midazolam | 1.22 | 326.1/291.2 | 326.1/244.2 | 326.1/209.0 |
14. | Triazolam | 2.05 | 343.1/315.0 | 343.1/308.2 | 343.1/239.2 |
15. | Temazepam | 2.12 | 301.2/283.1 | 301.2/255.1 | 301.2/228.1 |
16. | Flurazepam | 1.30 | 388.1/315.1 | 388.1/288.2 | 388.1/317.1 |
17. | Flunitrazepam | 2.23 | 314.4/211.0 | 314.4/239.1 | 314.4/268.1 |
18. | Clobazam | 2.26 | 301.1/259.1 | 301.1/224.0 | 301.1/181.1 |
19. | Diazepam-D5 (IS) | 2.40 | 290.0/154.1 | -- | -- |
20. | Diazepam | 2.44 | 284.8/193.1 | 284.8/222.2 | 284.8/257.1 |
Column | Ultra Biphenyl (cat.# 9109552) | ||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Dimensions: | 50 mm x 2.1 mm ID | ||||||||||||||||||||||||||||
Particle Size: | 5 µm | ||||||||||||||||||||||||||||
Pore Size: | 100 Å | ||||||||||||||||||||||||||||
Temp.: | 40 °C | ||||||||||||||||||||||||||||
Standard/Sample | |||||||||||||||||||||||||||||
Diluent: | Starting mobile phase + 30 ng/mL IS | ||||||||||||||||||||||||||||
Conc.: | 25 ng/mL sample was diluted 20x prior to injection | ||||||||||||||||||||||||||||
Inj. Vol.: | 30 µL | ||||||||||||||||||||||||||||
Mobile Phase | |||||||||||||||||||||||||||||
A: | Water + 2 mM ammonium formate + 0.2% formic acid | ||||||||||||||||||||||||||||
B: | Acetonitrile + 10% water + 2 mM ammonium formate + 0.2% formic acid | ||||||||||||||||||||||||||||
|
Detector | AB SCIEX API 4000 MS/MS |
---|---|
Model #: | API 4000 |
Ion Source: | TurboIonSpray® |
Ion Mode: | ESI+ |
Ion Spray Voltage: | 2 kV |
Curtain Gas: | 40 psi (275.8 kPa) |
Gas 1: | 40 psi (275.8 kPa) |
Gas 2: | 40 psi (275.8 kPa) |
Source Temp.: | 600 °C |
Mode: | MRM |
Instrument | Applied Biosystems/MDS Sciex LC-MS/MS System |
Notes | MS/MS instrument was operated in scheduled MRM mode. |