Peaks | tR (min) | Precursor Ion | Product Ion | Product Ion | |
---|---|---|---|---|---|
1. | Noroxymorphone | 1.35 | 288.1 | 270.1 | 212.9 |
2. | Morphine | 1.50 | 286.2 | 152.1 | 165.0 |
3. | Oxymorphone | 1.57 | 302.1 | 227.2 | 198.2 |
4. | Hydromorphone | 1.67 | 286.2 | 184.9 | 156.9 |
5. | Norcodeine | 1.80 | 286.1 | 151.9 | 115.0 |
6. | (-)-Dihydrocodeine | 1.91 | 302.2 | 199.0 | 128.2 |
7. | (-)-Naloxone | 1.92 | 328.3 | 310.1 | 212.3 |
8. | Codeine | 1.97 | 300.2 | 152.0 | 165.1 |
9. | O-Desmethyl-cis-tramadol | 2.01 | 250.1 | 58.0 | 42.0 |
10. | Noroxycodone | 2.04 | 302.1 | 227.0 | 197.9 |
11. | Desomorphine | 2.09 | 272.1 | 152.1 | 165.0 |
12. | 6-Acetylmorphine | 2.10 | 328.2 | 211.0 | 165.0 |
13. | 6-β-Naltrexol | 2.12 | 344.3 | 326.1 | 308.1 |
14. | Norhydrocodone | 2.13 | 286.1 | 199.0 | 128.2 |
15. | Naltrexone | 2.14 | 342.2 | 324.0 | 267.0 |
16. | Oxycodone | 2.15 | 316.2 | 298.0 | 169.0 |
17. | Tapentadol- β-D-glucuronide | 2.18 | 398.3 | 222.1 | 107.0 |
18. | Hydrocodone | 2.23 | 300.1 | 199.0 | 128.0 |
19. | Norbuprenorphine glucuronide | 2.26 | 590.5 | 414.3 | 165.0 |
20. | Norfentanyl | 2.31 | 233.1 | 84.1 | 55.0 |
21. | Nalbuphine | 2.33 | 358.2 | 340.1 | 115.0 |
Peaks | tR (min) | Precursor Ion | Product Ion | Product Ion | |
---|---|---|---|---|---|
22. | N-Desmethyltapentadol | 2.47 | 208.1 | 121.2 | 107.1 |
23. | Levorphanol | 2.52 | 258.2 | 156.8 | 199.2 |
24. | N-Desmethyl-cis-tramadol | 2.53 | 250.2 | 44.0 | 232.1 |
25. | Tramadol | 2.53 | 264.2 | 58.0 | 77.1 |
26. | Tapentadol | 2.55 | 222.1 | 121.0 | 107.1 |
27. | Heroin | 2.69 | 370.4 | 164.9 | 328.2 |
28. | Normeperidine | 2.70 | 234.1 | 160.2 | 91.0 |
29. | Meperidine | 2.75 | 248.1 | 220.1 | 174.1 |
30. | Buprenorphine glucuronide | 2.78 | 644.3 | 468.2 | 55.1 |
31. | Norbuprenorphine | 2.80 | 414.3 | 152.2 | 165.2 |
32. | Butorphanol | 2.87 | 328.3 | 310.3 | 131.2 |
33. | Pentazocine | 2.89 | 286.2 | 218.1 | 69.2 |
34. | Acetyl fentanyl | 2.91 | 323.3 | 105.0 | 188.0 |
35. | Fentanyl | 3.01 | 337.3 | 188.0 | 105.1 |
36. | Buprenorphine | 3.03 | 468.3 | 55.1 | 414.2 |
37. | Norpropoxyphene | 3.10 | 326.4 | 252.0 | 91.1 |
38. | Sufentanil | 3.12 | 387.2 | 238.1 | 111.1 |
39. | Propoxyphene | 3.13 | 340.3 | 265.9 | 58.1 |
40. | EMDP | 3.13 | 264.2 | 235.2 | 220.3 |
41. | Mitragynine | 3.14 | 399.3 | 174.2 | 159.0 |
42. | EDDP | 3.15 | 278.2 | 234.3 | 249.2 |
43. | Methadone | 3.18 | 310.2 | 264.9 | 105.1 |
Column | Raptor Biphenyl (cat.# 9309A12) | ||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Dimensions: | 100 mm x 2.1 mm ID | ||||||||||||||||||||||||
Particle Size: | 2.7 µm | ||||||||||||||||||||||||
Guard Column: | Raptor Biphenyl EXP guard column cartridge 5 mm, 2.1 mm ID, 2.7 µm (cat.# 9309A0252) | ||||||||||||||||||||||||
Temp.: | 30 °C | ||||||||||||||||||||||||
Standard/Sample | |||||||||||||||||||||||||
Diluent: | Water | ||||||||||||||||||||||||
Conc.: | 25 ng/mL | ||||||||||||||||||||||||
Inj. Vol.: | 5 µL | ||||||||||||||||||||||||
Mobile Phase | |||||||||||||||||||||||||
A: | Water + 0.1% formic acid | ||||||||||||||||||||||||
B: | Acetonitrile + 0.1% formic acid | ||||||||||||||||||||||||
|
Detector | MS/MS |
---|---|
Ion Mode: | ESI+ |
Mode: | Scheduled MRM |
Instrument | UHPLC |