Peaks | tR (min) | Precursor Ion | Quantifier Product Ion | Qualifier Product Ion | |
---|---|---|---|---|---|
1. | Pravadoline | 2.15 | 379.29 | 135.04 | 114.16 |
2. | AM2233 | 2.44 | 459.25 | 112.2 | 98.15 |
3. | JWH-200-d5 | 2.47 | 390.34 | 155.07 | NA |
4. | JWH-200 | 2.48 | 385.28 | 155.07 | 114.16 |
5. | WIN 55, 212 | 3.34 | 427.29 | 155.07 | 127.14 |
6. | JWH-073 N-butanoic acid | 3.39 | 358.27 | 155.08 | 127.11 |
7. | JWH-073 4-hydroxybutyl | 3.4 | 344.24 | 155.09 | 127.09 |
8. | JWH-018 N-pentanoic acid | 3.49 | 372.18 | 155.08 | 127.14 |
9. | JWH-018 5-hydroxypentyl-d5 | 3.54 | 363.5 | 155.08 | NA |
10. | JWH-018 5-hydroxypentyl | 3.55 | 358.27 | 155.08 | 127.11 |
11. | JWH-073 6-hydroxyindole | 3.77 | 344.24 | 155.09 | 127.09 |
12. | JWH-073 5-hydroxyindole-d7 | 3.81 | 351.21 | 155.07 | NA |
13. | JWH-073 5-hydroxyindole | 3.83 | 344.24 | 155.09 | 127.09 |
14. | JWH-073 7-hydroxyindole | 3.92 | 344.24 | 155.09 | 127.09 |
15. | JWH-018 6-hydroxyindole | 3.94 | 358.27 | 155.08 | 127.11 |
16. | JWH-018 5-hydroxyindole | 3.99 | 358.27 | 155.08 | 127.11 |
17. | JWH-018 7-hydroxyindole | 4.08 | 358.27 | 155.08 | 127.11 |
Peaks | tR (min) | Precursor Ion | Quantifier Product Ion | Qualifier Product Ion | |
---|---|---|---|---|---|
18. | RCS-4 | 4.15 | 322.27 | 135.12 | 77.09 |
19. | XLR-11 | 4.21 | 330.25 | 232.17 | 125.1 |
20. | JWH-015-d7 | 4.27 | 335.28 | 155.07 | NA |
21. | JWH-250 | 4.27 | 336.28 | 121.12 | 91.07 |
22. | JWH-015 | 4.29 | 328.26 | 155.07 | 127.13 |
23. | AM2201 | 4.30 | 360.26 | 155.07 | 127.14 |
24. | JWH-203 | 4.39 | 340.23 | 188.18 | 125.09 |
25. | JWH-073 | 4.42 | 328.26 | 155.07 | 127.13 |
26. | UR-144 | 4.44 | 312.32 | 214.17 | 125.1 |
27. | JWH-073 4-hydroxyindole | 4.53 | 344.24 | 155.09 | 127.09 |
28. | JWH-018-d9 | 4.55 | 351.34 | 155.07 | NA |
29. | JWH-018 | 4.57 | 342.27 | 155.08 | 127.11 |
30. | JWH-081 | 4.64 | 372.28 | 185.12 | 157.09 |
31. | JWH-018 4-hydroxyindole | 4.66 | 358.27 | 155.08 | 127.11 |
32. | JWH-122 | 4.69 | 356.29 | 169.12 | 141.11 |
33. | JWH-019 | 4.70 | 356.29 | 155.07 | 127.1 |
34. | JWH-210 | 4.84 | 370.31 | 183.12 | 153.26 |
Column | Raptor Biphenyl (cat.# 9309A5E) | ||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Dimensions: | 50 mm x 3.0 mm ID | ||||||||||||||||||||||||||||
Particle Size: | 2.7 µm | ||||||||||||||||||||||||||||
Guard Column: | Raptor Biphenyl EXP guard column cartridge 5.0 mm, 3.0 mm ID, 2.7 µm (cat.# 9309A0253) | ||||||||||||||||||||||||||||
Temp.: | 30 °C | ||||||||||||||||||||||||||||
Standard/Sample | |||||||||||||||||||||||||||||
Conc.: | 5 ng/mL standard was prepared in urine and diluted 3x with 50:50 methanol:water | ||||||||||||||||||||||||||||
Inj. Vol.: | 2 µL | ||||||||||||||||||||||||||||
Mobile Phase | |||||||||||||||||||||||||||||
A: | 0.1% Formic acid in water | ||||||||||||||||||||||||||||
B: | 0.1% Formic acid in acetonitrile | ||||||||||||||||||||||||||||
|
Detector | MS/MS |
---|---|
Ion Mode: | ESI+ |
Mode: | MRM |
Instrument | UHPLC |