Peaks | tR (min) | Conc. (µg/mL) | Precursor Ion | Product Ion | |
---|---|---|---|---|---|
1. | 4-Hydroxy-3-methoxymandelic acid-D3 (VMA-D3) | 0.90 | 0.83 | 200.1 | 139.9 |
2. | Vanillylmandelic acid (VMA) | 0.92 | 2.12–3.90* | 197.0 | 137.9 |
3. | 5-Hydroxyindole-4,6,7-D3-3-acetic-D2 acid (5-HIAA-D5) | 1.80 | 0.83 | 195.1 | 148.0 |
4. | 5-Hydroxyindole-3-acetic acid (5-HIAA) | 1.82 | 2.69–4.49* | 190.0 | 145.9 |
5. | 4-Hydroxy-3-methoxyphenyl-D3-acetic-D2 acid (HVA-D5) | 2.10 | 0.83 | 186.1 | 127.0 |
6. | Homovanillic acid (HVA) | 2.12 | 1.33–2.46* | 181.0 | 121.9 |
Column | Raptor Biphenyl (cat.# 9309512) | ||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Dimensions: | 100 mm x 2.1 mm ID | ||||||||||||||||||||
Particle Size: | 5 µm | ||||||||||||||||||||
Pore Size: | 90 Å | ||||||||||||||||||||
Guard Column: | Raptor Biphenyl guard column cartridge 5 mm, 2.1 mm ID, 5 µm (cat.# 930950252) | ||||||||||||||||||||
Temp.: | 30 °C | ||||||||||||||||||||
Standard/Sample | |||||||||||||||||||||
Diluent: | Water | ||||||||||||||||||||
Inj. Vol.: | 5 µL | ||||||||||||||||||||
Mobile Phase | |||||||||||||||||||||
A: | 0.1% Formic acid, 5 mM ammonium formate in water | ||||||||||||||||||||
B: | Methanol | ||||||||||||||||||||
|
Detector | MS/MS |
---|---|
Ion Mode: | ESI- |
Mode: | MRM |
Instrument | UHPLC |
Sample Preparation | The human urine sample is the Bio-Rad Lyphochek quantitative urine control, level 1. A 40 μL aliquot of urine was mixed with 360 μL of water and 10 μL of internal standard solution (33.3 μg/mL in methanol) in a Thomson 0.45 μm PVDF filter vial and injected for analysis after filtration. |