Peaks | tR (min) | Q1 | Q3 Quantifier | Q3 Qualifier | |
---|---|---|---|---|---|
1. | Cortisol | 2.3 | 363.3 | 121.2 | 327.2 |
2. | Estradiol | 2.6 | 255.3 | 159.1 | 133.1 |
3. | 11-Deoxycortisol | 4.0 | 347.4 | 109.2 | 97.1 |
4. | DHEA | 4.3 | 271.3 | 253.2 | 213.2 |
5. | Boldenone | 4.4 | 287.2 | 121.2 | 135.2 |
6. | Corticosterone | 4.4 | 347.4 | 329.2 | 121.1 |
7. | Stanozolol | 4.4 | 329.3 | 81.1 | 95.1 |
8. | Epitestosterone | 4.5 | 289.2 | 109.2 | 97.1 |
Peaks | tR (min) | Q1 | Q3 Quantifier | Q3 Qualifier | |
---|---|---|---|---|---|
9. | 17-OH-progesterone | 4.8 | 331.3 | 97.2 | 109.5 |
10. | Testosterone | 5.0 | 289.2 | 109.1 | 97.0 |
11. | Trenbolone | 5.1 | 303.3 | 109.2 | 97.1 |
12. | Methyltestosterone | 5.1 | 271.2 | 253.1 | 199.2 |
13. | Nandrolone | 5.3 | 275.2 | 109.2 | 257.3 |
14. | 25-Hydroxyvitamin D3 | 6.1 | 383.3 | 365.3 | 107.1 |
15. | Androstenedione | 6.2 | 287.2 | 97.1 | 109.2 |
16. | Progesterone | 7.5 | 315.3 | 109.2 | 97.1 |
Column | Force Biphenyl LC Column (cat.# 9629512) | ||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Dimensions: | 100 mm x 2.1 mm ID | ||||||||||||||||||||
Particle Size: | 5 µm | ||||||||||||||||||||
Pore Size: | 100 Å | ||||||||||||||||||||
Temp.: | 30 °C | ||||||||||||||||||||
Standard/Sample | Steroids panel | ||||||||||||||||||||
Diluent: | 50:50 Methanol:water | ||||||||||||||||||||
Conc.: | 5 ng/mL | ||||||||||||||||||||
Inj. Vol.: | 50 µL | ||||||||||||||||||||
Mobile Phase | |||||||||||||||||||||
A: | Water, 0.1% formic acid | ||||||||||||||||||||
B: | Methanol, 0.1% formic acid | ||||||||||||||||||||
| |||||||||||||||||||||
Max Pressure: | 250 bar |
Detector | MS/MS |
---|---|
Ion Mode: | ESI+ |
Mode: | Scheduled MRM |
Instrument | UHPLC |
Notes |