Peaks | tR (min) | Precursor Ion | Product Ion | |
---|---|---|---|---|
1. | Norepinephrine-d6 | 1.14 | 158.16 | 111.26 |
2. | Norepinephrine | 1.14 | 152.07 | 106.91 |
3. | Epinephrine-d6 | 1.32 | 172.16 | 111.86 |
4. | Epinephrine | 1.33 | 166.07 | 135.00 |
5. | Normetanephrine-d3 | 1.51 | 169.00 | 136.96 |
6. | Normetanephrine | 1.52 | 166.00 | 148.99 |
Peaks | tR (min) | Precursor Ion | Product Ion | |
---|---|---|---|---|
7. | Dopamine-d4 | 1.55 | 141.16 | 94.69 |
8. | Dopamine | 1.56 | 136.97 | 64.66 |
9. | Metanephrine-d3 | 1.94 | 183.00 | 151.15 |
10. | Metanephrine | 1.96 | 179.94 | 165.01 |
11. | 3-Methoxytyramine-d4 | 2.42 | 155.07 | 122.93 |
12. | 3-Methoxytyramine | 2.45 | 151.00 | 91.02 |
Column | Raptor Biphenyl (cat.# 9309A62) | ||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Dimensions: | 150 mm x 2.1 mm ID | ||||||||||||||||||||||||
Particle Size: | 2.7 µm | ||||||||||||||||||||||||
Pore Size: | 90 Å | ||||||||||||||||||||||||
Temp.: | 30 °C | ||||||||||||||||||||||||
Standard/Sample | |||||||||||||||||||||||||
Conc.: | Endogenous level | ||||||||||||||||||||||||
Inj. Vol.: | 2 µL | ||||||||||||||||||||||||
Mobile Phase | |||||||||||||||||||||||||
A: | Water, 0.2% formic acid | ||||||||||||||||||||||||
B: | Methanol, 0.2% formic acid | ||||||||||||||||||||||||
|
Detector | MS/MS |
---|---|
Ion Mode: | ESI+ |
Mode: | MRM |
Instrument | UHPLC |
Sample Preparation | A 200 µL aliquot of urine sample (Bio-Rad Lyphochek quantitative urine control, normal level) was mixed with 10 µL (1 µg/mL in methanol) internal standard solution and 600 μL of 250 mM ammonium acetate solution. The mixture was loaded onto the EVOLUTE EXPRESS WCX 96-well plate (30 mg), washed with 1 mL water and 1 mL methanol:acetonitrile (60:40), eluted with 200 µL of 5% formic acid in water:methanol (95:5), and injected (2 μL) for analysis. |