Peaks | tR (min) | Conc. (ng/mL) | Precursor Ion | Product Ion | Product Ion | |
---|---|---|---|---|---|---|
1. | Ethyl-β-D-glucuronide-d5 | 0.77 | 100 | 225.9 | 84.7 | - |
2. | Ethyl-β-D-glucuronide | 0.78 | 500 | 220.9 | 74.9 | 85.0 |
3. | Ethyl sulfate-d5 | 1.75 | 100 | 129.8 | 97.7 | - |
4. | Ethyl sulfate | 1.78 | 500 | 124.8 | 96.8 | 79.7 |
Column | Raptor EtG/EtS (cat.# 9325A12) | ||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Dimensions: | 100 mm x 2.1 mm ID | ||||||||||||||||||||
Particle Size: | 2.7 µm | ||||||||||||||||||||
Pore Size: | 90 Å | ||||||||||||||||||||
Guard Column: | UltraShield UHPLC precolumn filter (cat.# 25809) | ||||||||||||||||||||
Temp.: | 30 °C | ||||||||||||||||||||
Standard/Sample | Ethyl- | ||||||||||||||||||||
Ethyl- | |||||||||||||||||||||
Ethyl sulfate sodium salt (EtS) (cat.# 34103) | |||||||||||||||||||||
Ethyl sulfate-d5 sodium salt (EtS-d5) (cat.# 34104) | |||||||||||||||||||||
Diluent: | 0.1% Formic acid in water | ||||||||||||||||||||
Conc.: | 500 ng/mL | ||||||||||||||||||||
Inj. Vol.: | 10 µL | ||||||||||||||||||||
Mobile Phase | |||||||||||||||||||||
A: | 0.1% Formic acid in water | ||||||||||||||||||||
B: | 0.1% Formic acid in acetonitrile | ||||||||||||||||||||
|
Detector | MS/MS |
---|---|
Ion Mode: | ESI- |
Mode: | MRM |
Instrument | HPLC |
Sample Preparation | A 500 ng/mL standard was prepared in oral fluid. 50 µL of the standard was diluted with 950 µL of a working internal standard (100 ng/mL EtG-d5 and EtS-d5 in 0.1% formic acid in water). The sample was vortexed at 3000 rpm for 10 seconds to mix. The sample was then centrifuged at 3000 rpm for 10 minutes at 10 °C. The autosampler needle was adjusted to inject from the supernatant. |