Peaks | tR (min) | Precursor ion | Product ion | |
---|---|---|---|---|
1. | γ-Hydroxybutyric acid (GHB) | 1.42 | 105.2 | 87.0 |
2. | 1,4-Butanediol (1,4-BD) | 1.42 | 91.0 | 55.0 |
3. | γ-Butyrolactone (GBL) | 1.75 | 87.0 | 45.0 |
Column | Force C18 (cat.# 963431E) | ||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Dimensions: | 100 mm x 3.0 mm ID | ||||||||||||||||||||||||
Particle Size: | 3 µm | ||||||||||||||||||||||||
Pore Size: | 100 Å | ||||||||||||||||||||||||
Guard Column: | Force C18 EXP guard column cartridge 5 mm, 3.0 mm ID, 3 µm (cat.# 963450253) | ||||||||||||||||||||||||
Temp.: | 30 °C | ||||||||||||||||||||||||
Standard/Sample | |||||||||||||||||||||||||
Diluent: | Water | ||||||||||||||||||||||||
Conc.: | 500 ng/mL | ||||||||||||||||||||||||
Inj. Vol.: | 10 µL | ||||||||||||||||||||||||
Mobile Phase | |||||||||||||||||||||||||
A: | 0.5% Formic acid in water | ||||||||||||||||||||||||
B: | 0.5% Formic acid in methanol | ||||||||||||||||||||||||
|
Detector | MS/MS |
---|---|
Ion Mode: | ESI+ |
Mode: | MRM |
Instrument | HPLC |
Sample Preparation | 100 μL of whole human blood was fortified at 50 μg/mL with GHB, GBL, 1,4-BD, and GHB-D6 (IS) using 5 μL of 1 mg/mL solutions. The blood was precipitated with 380 μL methanol. The sample was then vortexed at 1000 rpm for 10 seconds and centrifuged at 3000 rpm for 10 minutes at 10 °C. 50 μL of the supernatant was removed and diluted to 1 mL with water. The sample was then vortexed and subjected to LC-MS/MS analysis. (Internal standard not shown on chromatogram.) |