Peaks | tR (min) | Conc. (ng/mL) | Precursor Ion | Product Ion 1 | Product Ion 2 | |
---|---|---|---|---|---|---|
1. | Psilocin | 1.13 | 100 | 205.1 | 160.1 | 115.0 |
2. | Psilocybin | 1.75 | 100 | 285.1 | 240.0 | 205.2 |
Column | Raptor HILIC-Si (cat.# 9310A52) | ||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Dimensions: | 50 mm x 2.1 mm ID | ||||||||||||||||||||||||
Particle Size: | 2.7 µm | ||||||||||||||||||||||||
Pore Size: | 90 Å | ||||||||||||||||||||||||
Temp.: | 40 °C | ||||||||||||||||||||||||
Standard/Sample | |||||||||||||||||||||||||
Diluent: | 10 mM Ammonium formate in 100% methanol | ||||||||||||||||||||||||
Conc.: | 100 ng/mL | ||||||||||||||||||||||||
Inj. Vol.: | 2 µL | ||||||||||||||||||||||||
Mobile Phase | |||||||||||||||||||||||||
A: | Water, 10 mM ammonium formate | ||||||||||||||||||||||||
B: | 90:10 Acetonitrile:water (v/v), 10 mM ammonium formate | ||||||||||||||||||||||||
|
Detector | MS/MS |
---|---|
Ion Source: | Electrospray |
Ion Mode: | ESI+ |
Mode: | MRM |
Sample Preparation | A 100 ng/mL standard mix of psilocin and psilocybin was prepared in 10 mM ammonium formate in 100% methanol. The solution was vortexed at 3000 rpm for 10 seconds to mix, and then the supernatant was injected for LC-MS/MS analysis. |