Peaks | tR (min) | Conc. (ng/mL) | Precursor Ion | Product Ion 1 | Product Ion 2 | |
---|---|---|---|---|---|---|
1. | Psilocybin | 2.09 | 500 | 285.1 | 205.1 | 240.0 |
2. | Psilocin-D10 | 2.74 | 200 | 215.1 | 66.1 | - |
3. | Psilocin | 2.78 | 500 | 205.1 | 160.1 | 115.0 |
Column | Raptor Biphenyl (cat.# 9309A12) | ||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Dimensions: | 100 mm x 2.1 mm ID | ||||||||||||||||||||||||
Particle Size: | 2.7 µm | ||||||||||||||||||||||||
Pore Size: | 90 Å | ||||||||||||||||||||||||
Guard Column: | Raptor Biphenyl EXP guard column cartridge 5.0 mm, 2.1 mm ID, 2.7 µm (cat.# 9309A0252) | ||||||||||||||||||||||||
Temp.: | 30 °C | ||||||||||||||||||||||||
Standard/Sample | |||||||||||||||||||||||||
Diluent: | Water, 0.1% formic acid + 2 mM ammonium formate | ||||||||||||||||||||||||
Conc.: | 500 ng/mL | ||||||||||||||||||||||||
Inj. Vol.: | 5 µL | ||||||||||||||||||||||||
Mobile Phase | |||||||||||||||||||||||||
A: | Water, 0.1% formic acid + 2 mM ammonium formate | ||||||||||||||||||||||||
B: | Methanol, 0.1% formic acid + 2 mM ammonium formate | ||||||||||||||||||||||||
|
Detector | MS/MS |
---|---|
Ion Source: | Electrospray |
Ion Mode: | ESI+ |
Mode: | MRM |
Instrument | HPLC |
Sample Preparation | A 500 ng/mL standard mix of psilocin and psilocybin was prepared in pooled urine. A 50 µL aliquot was taken from the standard and mixed with 10 µL of internal standard (psilocin-D10, 20 µg/mL) and 100 µL of methanol. The mixture was vortexed at 3000 rpm for 10 seconds and centrifuged at 4300 rpm for 10 minutes at 10 °C. After centrifugation, 100 µL of the supernatant was diluted with 900 µL (20-fold dilution) of water containing 0.1% formic acid and 2 mM ammonium formate (mobile phase A) and injected for LC-MS/MS analysis. |