Peaks | |
---|---|
1. | HMX |
2. | RDX |
3. | 1,3,5-Trinitrobenzene |
4. | 1,4-Dinitrobenzene |
5. | 1,3-Dinitrobenzene |
6. | 3,5-Dinitroaniline |
7. | Nitrobenzene |
8. | Nitroglycerin |
9. | Tetryl |
Peaks | |
---|---|
10. | 2,4,6-Trinitrotoluene |
11. | 2-Amino-4,6-dinitrotoluene |
12. | 4-Amino-2,6-dinitrotoluene |
13. | 2,4-Dinitrotoluene |
14. | 2,6-Dinitrotoluene |
15. | 2-Nitrotoluene |
16. | 4-Nitrotoluene |
17. | 3-Nitrotoluene |
18. | PETN |
Column | Raptor ARC-18 (cat.# 9314A1E) | ||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Dimensions: | 100 mm x 3.0 mm ID | ||||||||||||
Particle Size: | 2.7 µm | ||||||||||||
Temp.: | 35 °C | ||||||||||||
Standard/Sample | 8330B Nitroaromatics and nitramine mix (cat.# 33204) | ||||||||||||
1,4-Dinitrobenzene (cat.# 33205) | |||||||||||||
Diluent: | Water | ||||||||||||
Conc.: | 20 µg/mL | ||||||||||||
Inj. Vol.: | 3 µL | ||||||||||||
Mobile Phase | |||||||||||||
A: | Water | ||||||||||||
B: | Methanol | ||||||||||||
|
Detector | UV/Vis @ 210/254 nm |
---|---|
Cell Temp: | 35 °C |
Flow Cell Size: | Semi-micro |
Instrument | Shimadzu UFLCXR |
Notes | This is the recommended confirmation column for a two-column analysis. See LC_EV0531 for the primary column. |