Perfluorinated Compounds (PFCs) on Raptor C18 (5 μm)

LC_EV0541
PeakstR (min)Conc.
(ng/mL)
Precursor IonProduct Ion
1.Perfluorobutanoic acid (PFBA)0.845213.07169.07
2.Perfluoropentanoic acid (PFPeA)1.765263.10219.11
3.Perfluorobutane sulfonic acid (PFBS)2.035299.1079.99
4.1H, 1H, 2H, 2H-perfluorohexane sulfonic acid (4:2 FTS)2.585327.10307.08
5.Perfluorohexanoic acid (PFHxA)2.655313.10269.12
6.Perfluoropentane sulfonic acid (PFPeS)2.795349.1079.98
7.Perfluoroheptanoic acid (PFHpA)3.295363.16319.09
8.Perfluorohexane sulfonic acid (PFHxS)3.355399.1379.98
9.2-Perfluorohexyl ethanoic acid (FHEA)3.41250377.13293.07
10.1H, 1H, 2H, 2H-perfluorooctane sulfonic acid (6:2 FTS)3.735427.17407.18
11.Perfluorooctanoic acid (PFOA)3.755413.16369.10
12.Perfluoroheptane sulfonic acid (PFHpS)3.795449.1779.98
13.Perfluorononanoic acid (PFNA)4.145463.16419.19
14.Perfluorooctane sulfonic acid (PFOS)4.165499.1779.98
15.2-Perfluorooctyl ethanoic acid (FOEA)4.26250477.17393.15
16.1H, 1H, 2H, 2H-perfluorodecane sulfonic acid (8:2 FTS)4.465527.17507.16
17.Perfluorodecanoic acid (PFDA)4.475513.17469.16
PeakstR (min)Conc.
(ng/mL)
Precursor IonProduct Ion
18.Perfluorononane sulfonic acid (PFNS)4.475549.1779.98
19.N-methyl perfluorooctane sulfonamidoacetic acid (N-MeFOSAA)4.605570.20419.17
20.Perfluorodecane sulfonic acid (PFDS)4.745599.1779.98
21.N-ethyl perfluorooctane sulfonamidoacetic acid (N-EtFOSAA)4.745584.20419.18
22.Perfluorooctane sulfonamide (FOSA)4.745498.1777.97
23.Perfluoroundecanoic acid (PFUnA)4.745563.23519.24
24.2-Perfluorodecyl ethanoic acid (FDEA)4.85250577.23493.18
25.Perfluorododecanoic acid (PFDoA)4.995613.23569.19
26.1H, 1H, 2H, 2H-perfluorododecane sulfonic acid (10:2 FTS)5.005627.23607.20
27.Perfluorotridecanoic acid (PFTrDA)5.205663.23619.21
28.N-methyl perfluorooctane sulfonamide (N-MeFOSA)5.245512.17169.07
29.N-methyl perfluorooctane sulfonamidoethanol (N-MeFOSE)5.2550616.2759.11
30.Perfluorotetradecanoic acid (PFTeDA)5.395713.23669.23
31.N-ethyl perfluorooctane sulfonamide (N-EtFOSA)5.415526.23169.06
32.N-ethyl perfluorooctane sulfonamidoethanol (N-EtFOSE)5.4150630.3359.11
33.Perfluorohexadecanoic acid (PFHxDA)5.695813.30769.28
34.Perfluorooctadecanoic acid (PFODA)5.925913.30869.32
ColumnRaptor C18 (cat.# 9304552)
Dimensions:50 mm x 2.1 mm ID
Particle Size:5 µm
Pore Size:90 Å
Temp.:40 °C
Standard/Sample
Diluent:80:20 Water:methanol
Conc.:5-250 ng/mL
Inj. Vol.:5 µL
Mobile Phase
A:5 mM Ammonium acetate in water
B:Methanol
Time (min)Flow (mL/min)%A%B
0.000.48020
6.000.4595
6.500.4595
6.510.48020
8.000.48020
DetectorMS/MS
Ion Mode:ESI-
Mode:MRM
InstrumentUHPLC