Peaks | tR (min) | Conc. (ng/mL) | Precursor Ion | Product Ion | |
---|---|---|---|---|---|
1. | Perfluorobutanesulfonic acid (PFBS) | 1.98 | 10 | 299.10 | 79.99 |
2. | Perfluoro-n-[1,2-13C2]hexanoic acid (13C-PFHxA) | 2.92 | 10 | 315.13 | 270.13 |
3. | Perfluorohexanoic acid (PFHxA) | 2.93 | 10 | 313.10 | 269.12 |
4. | Perfluoroheptanoic acid (PFHpA) | 4.00 | 5 | 363.16 | 319.09 |
5. | Perfluorohexanesulfonic acid (PFHxS) | 4.14 | 10 | 399.13 | 79.98 |
6. | Perfluoro-[1,2-13C2]octanoic acid (13C-PFOA) | 4.83 | 5 | 415.13 | 370.10 |
7. | Perfluorooctanoic acid (PFOA) | 4.83 | 5 | 413.16 | 369.10 |
8. | Perfluorononanoic acid (PFNA) | 5.50 | 5 | 463.16 | 419.19 |
9. | Perfluoro-1-[1,2,3,4-13C4]octanesulfonic acid (13C-PFOS) | 5.54 | 5 | 503.13 | 79.98 |
10. | Perfluorooctanesulfonic acid (PFOS) | 5.54 | 10 | 499.17 | 79.98 |
Peaks | tR (min) | Conc. (ng/mL) | Precursor Ion | Product Ion | |
---|---|---|---|---|---|
11. | Perfluoro-n-[1,2-13C2]decanoic acid (13C-PFDA) | 6.06 | 5 | 515.17 | 470.17 |
12. | Perfluorodecanoic acid (PFDA) | 6.07 | 5 | 513.17 | 469.16 |
13. | N-deuteriomethylperfluoro-1-octanesulfonamidoacetic acid (d3-NMeFOSAA) | 6.28 | 5 | 573.23 | 419.15 |
14. | N-methyl perfluorooctanesulfonamidoacetic acid (NMeFOSAA) | 6.28 | 5 | 570.20 | 419.17 |
15. | N-deuterioethylperfluoro-1-octanesulfonamidoacetic acid (d5-NEtFOSAA) | 6.52 | 5 | 589.27 | 419.11 |
16. | N-ethyl perfluorooctanesulfonamidoacetic acid (NEtFOSAA) | 6.52 | 5 | 584.20 | 419.18 |
17. | Perfluoroundecanoic acid (PFUnA) | 6.56 | 5 | 563.23 | 519.24 |
18. | Perfluorododecanoic acid (PFDoA) | 6.98 | 5 | 613.23 | 569.19 |
19. | Perfluorotridecanoic acid (PFTrDA) | 7.35 | 5 | 663.23 | 619.21 |
20. | Perfluorotetradecanoic acid (PFTA) | 7.67 | 5 | 713.23 | 669.23 |
Column | Raptor C18 (cat.# 9304552) | ||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Dimensions: | 50 mm x 2.1 mm ID | ||||||||||||||||||||
Particle Size: | 5 µm | ||||||||||||||||||||
Pore Size: | 90 Å | ||||||||||||||||||||
Temp.: | 40 °C | ||||||||||||||||||||
Standard/Sample | |||||||||||||||||||||
Diluent: | 96:4 Methanol:water | ||||||||||||||||||||
Conc.: | 5–10 ng/mL | ||||||||||||||||||||
Inj. Vol.: | 2 µL | ||||||||||||||||||||
Mobile Phase | |||||||||||||||||||||
A: | 5 mM ammonium acetate in water | ||||||||||||||||||||
B: | Methanol | ||||||||||||||||||||
|
Detector | MS/MS |
---|---|
Ion Source: | Electrospray |
Ion Mode: | ESI- |
Mode: | MRM |
Instrument | Waters ACQUITY |