Welcome to the New Restek.com! Take a tour of the new site and feel free to send us your feedback.

Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.

Perfluorinated Alkyl Acids (PFBS, PFOA, PFOS, and HFPO-DA/GenX) on Raptor C18 5 μm

LC_EV0553
PeakstR (min)Conc.
(ng/mL)
Precursor IonProduct Ion
1.Perfluorobutanesulfonic acid (PFBS)1.5610298.879.9
2.Hexafluoropropylene oxide-dimer acid (HFPO-DA/GenX)2.0210328.9284.9
3.Perfluorooctanoic acid (PFOA)2.475413.1368.9
4.Perfluorooctanesulfonic acid (PFOS)2.6510498.880.0
ColumnRaptor C18 (cat.# 9304552)
Dimensions:50 mm x 2.1 mm ID
Particle Size:5 µm
Pore Size:90 Å
Temp.:40 °C
Sample
Diluent:75:25 Water:methanol
Conc.:5-10 ng/mL
Inj. Vol.:5 µL
Mobile Phase
A:Water, 5 mM ammonium acetate
B:Methanol
Time (min)Flow (mL/min)%A%B
0.000.47525
3.000.4595
3.010.47525
4.000.47525
DetectorMS/MS
Ion Mode:ESI-
Mode:MRM
InstrumentUHPLC