Peaks | |
---|---|
1. | Benzo[c]fluorene |
2. | Benzo[a]anthracene |
3. | Chrysene |
4. | 5-Methylchrysene |
5. | Benzo(j)fluoranthene |
6. | Benzo[b]fluoranthene |
7. | Benzo[k]fluoranthene |
Peaks | |
---|---|
8. | Benzo[a]pyrene |
9. | Dibenzo[a,l]pyrene |
10. | Dibenz[a,h]anthracene |
11. | Benzo[ghi]perylene |
12. | Indeno[1,2,3-cd]pyrene |
13. | Dibenzo[a,e]pyrene |
14. | Dibenzo(a,i)pyrene |
15. | Dibenzo(a,h)pyrene |
Column | Pinnacle II PAH, custom (cat.# 921946E) | ||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Dimensions: | 150 mm x 3.0 mm ID | ||||||||||||||||||||||||||||
Particle Size: | 4 µm | ||||||||||||||||||||||||||||
Pore Size: | 110 Å | ||||||||||||||||||||||||||||
Temp.: | 30 °C | ||||||||||||||||||||||||||||
Standard/Sample | custom EU 15+1 PAH mix | ||||||||||||||||||||||||||||
Diluent: | acetonitrile | ||||||||||||||||||||||||||||
Conc.: | 30 ng/mL (ppb) | ||||||||||||||||||||||||||||
Inj. Vol.: | 5 µL | ||||||||||||||||||||||||||||
Mobile Phase | |||||||||||||||||||||||||||||
A: | water | ||||||||||||||||||||||||||||
B: | 95:5 acetonitrile:tetrahydrofuran | ||||||||||||||||||||||||||||
| |||||||||||||||||||||||||||||
Max Pressure: | 275 bar |
Detector | Shimadzu RF-10AXL | ||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| |||||||||||||||||||||||||||||
Instrument | Shimadzu Prominence | ||||||||||||||||||||||||||||
Notes | RF-10AXL sensitivity: medium Cyclopenta(c,d)pyrene was included in the EU 15+1 PAH mix but it does not fluoresce. |
We have determined your location as:
Or select your location below: