Peaks | |
---|---|
1. | Methamidophos |
2. | Omethoate |
3. | Thiamethoxam |
4. | Dimethoate |
5. | Clothianidin |
6. | Imidacloprid |
Peaks | |
---|---|
7. | Propamocarb |
8. | Carbaryl |
9. | Metalaxyl |
10. | Diphenylamine |
11. | Myclobutanil |
12. | Spinosyn A |
13. | Spinosyn D |
Column | Ultra Aqueous C18 (cat.# 9178312) |
---|---|
Dimensions: | 100 mm x 2.1 mm ID |
Particle Size: | 3 µm |
Pore Size: | 100 Å |
Standard/Sample | QuEChERS extract of red bell pepper |
Nicarbazin (bis-nitrophenol urea) (cat.# 33261) | |
Diluent: | Deionized water:acetonitrile (90:10) |
Inj. Vol.: | 10 µL |
Mobile Phase | 10 mM ammonium acetate in water:10 mM ammonium formate in methanol with gradient program |
Flow: | 0.5 mL/min |
Detector | AB SCIEX API 4000 LC/MS/MS System |
---|---|
Ion Source: | TurboIonSpray® |
Ion Mode: | ESI+ |
Mode: | MRM |
Instrument | API LC-MS/MS |
Sample Preparation | Internal Standard: 100 µL of QuEChERS internal standard mix for GC-MS analysis (cat.# 33267) and 100 µL of QuEChERS internal standard mix for LC-MS/MS analysis (cat.# 33261) were added to each 10 g sample. These internal standard mixes require no dilutions and contain compounds specified in the EN 15662 QuEChERS method. Extraction: 10 mL acetonitrile was added to a 10 gram sample of homogenized red bell pepper. After a 1 minute shake, Q-sep European method extraction salts (cat.# 26236) were added followed by another 1 minute shake. The sample was then centrifuged with a Q-sep 3000 centrifuge (cat.# 26230). dSPE: Add 1 mL extract to QuEChERS dSPE cleanup tube containing 150 mg MgSO4, 25 mg PSA, and 25 mg C18 (cat.# 26216). Shake for 30 seconds and centrifuge with a Q-sep 3000 centrifuge (cat.# 26230). |