Peaks | tR (min) | Precursor Ion (m/z) | Product Ion 1 (m/z) | Product Ion 2 (m/z) | |
---|---|---|---|---|---|
1. | Fuberidazole | 4.6 | 185.0 | 157.0 | 65.0 |
2. | Ethirimol | 5.0 | 210.2 | 140.1 | 98.1 |
3. | Flutriafol | 5.9 | 302.1 | 70.1 | 123.0 |
4. | Nuarimol | 6.8 | 315.0 | 252.1 | 81.0 |
5. | Paclobutrazol | 7.0 | 294.0 | 70.0 | 125.0 |
6. | Cyproconazole isomer 1 | 7.3 | 292.0 | 70.0 | 125.0 |
7. | Triadimenol | 7.3 | 296.1 | 70.0 | 227.1 |
8. | Cyproconazole isomer 2 | 7.7 | 292.0 | 70.0 | 125.0 |
9. | Bromuconazole isomer 1 | 7.8 | 378.0 | 159.0 | 70.0 |
10. | Fenarimol | 7.8 | 331.0 | 268.0 | 81.0 |
11. | Fluquinconazole | 7.8 | 376.0 | 307.0 | 349.0 |
12. | Tetraconazole | 7.9 | 372.1 | 159.0 | 70.0 |
13. | Triticonazole | 8.0 | 318.1 | 70.0 | 125.0 |
14. | Etaconazole isomer 1 | 8.3 | 328.1 | 159.0 | 205.0 |
15. | Fenbuconazole | 8.4 | 337.0 | 124.9 | 70.0 |
16. | Diclobutrazol | 8.5 | 328.2 | 70.2 | 59.1 |
17. | Epoxiconazole | 8.5 | 330.0 | 121.1 | 101.1 |
Peaks | tR (min) | Precursor Ion (m/z) | Product Ion 1 (m/z) | Product Ion 2 (m/z) | |
---|---|---|---|---|---|
18. | Etaconazole isomer 2 | 8.5 | 328.1 | 159.0 | 205.0 |
19. | Flusilazole | 8.5 | 316.1 | 165.1 | 247.1 |
20. | Tebuconazole | 8.9 | 308.2 | 70.0 | 125.0 |
21. | Bromuconazole isomer 2 | 9.2 | 378.0 | 159.0 | 70.0 |
22. | Hexaconazole | 9.2 | 314.1 | 70.0 | 159.0 |
23. | Penconazole | 9.3 | 284.1 | 70.0 | 159.0 |
24. | Bitertanol | 9.4 | 338.2 | 269.2 | 70.0 |
25. | Metconazole | 9.4 | 320.1 | 70.0 | 125.0 |
26. | Propiconazole isomer 1 | 9.4 | 342.1 | 159.0 | 69.0 |
27. | Diniconazole | 9.5 | 326.2 | 70.2 | 159.0 |
28. | Propiconazole isomer 2 | 9.5 | 342.1 | 159.0 | 69.0 |
29. | Ipconazole isomer 1 | 9.8 | 334.2 | 70.0 | 125.0 |
30. | Ipconazole isomer 2 | 10.0 | 334.2 | 70.0 | 125.0 |
31. | Difenoconazole isomer 1 | 10.1 | 408.2 | 253.1 | 251.1 |
32. | Difenoconazole isomer 2 | 10.2 | 408.2 | 253.1 | 251.1 |
33. | Triflumizole | 10.2 | 346.1 | 278.1 | 73.0 |
34. | Etoxazole | 11.0 | 360.1 | 141.0 | 57.2 |
Column | Ultra Aqueous C18 (cat.# 9178312) | ||||||||||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Dimensions: | 100 mm x 2.1 mm ID | ||||||||||||||||||||||||||||||||||||
Particle Size: | 3 µm | ||||||||||||||||||||||||||||||||||||
Pore Size: | 100 Å | ||||||||||||||||||||||||||||||||||||
Temp.: | 50 °C | ||||||||||||||||||||||||||||||||||||
Standard/Sample | LC multiresidue pesticide standard #6 (cat.# 31977) | ||||||||||||||||||||||||||||||||||||
Diluent: | Water | ||||||||||||||||||||||||||||||||||||
Conc.: | 50 ng/mL | ||||||||||||||||||||||||||||||||||||
Inj. Vol.: | 20 µL | ||||||||||||||||||||||||||||||||||||
Mobile Phase | |||||||||||||||||||||||||||||||||||||
A: | Water + 4 mM ammonium formate + 0.1% formic acid | ||||||||||||||||||||||||||||||||||||
B: | Methanol + 4 mM ammonium formate + 0.1% formic acid | ||||||||||||||||||||||||||||||||||||
| |||||||||||||||||||||||||||||||||||||
Max Pressure: | 255 bar |
Detector | AB SCIEX API 4000 LC-MS/MS |
---|---|
Ion Source: | TurboIonSpray® |
Ion Mode: | ESI+ |
Ion Spray Voltage: | 5.5 kV |
Curtain Gas: | 30 psi (206.8 kPa) |
Gas 1: | 40 psi (275.8 kPa) |
Gas 2: | 45 psi (310.3 kPa) |
CAD: | 10 psi (68.9 kPa) |
Source Temp.: | 350 °C |
Mode: | Scheduled MRM |
MRM Detection Window: | 60 sec |
Target Scan Time: | 0.33 sec |
Instrument | API LC-MS/MS |
Notes | Q1/Q3 Resolution: Unit Autosampler Temp.: 5 °C |