Peaks | tR (min) | Q1 | Q3 Quantifier | Q3 Qualifier | |
---|---|---|---|---|---|
1. | Aflatoxin G2 | 1.07 | 331.1 | 245.1 | 189.1 |
2. | Aflatoxin G1 | 1.13 | 329.0 | 243.1 | 200.1 |
3. | Aflatoxin B2 | 1.23 | 315.0 | 259.1 | 287.1 |
4. | Aflatoxin B1 | 1.29 | 313.0 | 285.1 | 241.1 |
Column | Raptor ARC-18 (cat.# 9314A5E) | ||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Dimensions: | 50 mm x 3.0 mm ID | ||||||||||||||||||||
Particle Size: | 2.7 µm | ||||||||||||||||||||
Temp.: | 45 °C | ||||||||||||||||||||
Standard/Sample | |||||||||||||||||||||
Diluent: | Acetonitrile:water (50:50) | ||||||||||||||||||||
Conc.: | 100 ng/mL | ||||||||||||||||||||
Inj. Vol.: | 10 µL | ||||||||||||||||||||
Mobile Phase | |||||||||||||||||||||
A: | 5 mM Ammonium formate + 0.1% formic acid in water | ||||||||||||||||||||
B: | 0.1% Formic acid in methanol | ||||||||||||||||||||
|
Detector | Applied Biosystems/MDS Sciex LC-MS/MS |
---|---|
Ion Source: | TurboIonSpray® |
Ion Mode: | ESI+ |
Instrument | Shimadzu UFLCXR |