Peaks | tR (min) | Conc. (µmole/mL) | |
---|---|---|---|
1. | Arginine (Arg) | 5.145 | 0.4 |
2. | Asparagine (Asn) | 5.518 | 0.4 |
3. | Glutamine (Gln) | 5.630 | 0.4 |
4. | trans-4-Hydroxyproline (Hyp) | 5.878 | 0.4 |
5. | Serine (Ser) | 6.139 | 0.4 |
6. | Aspartic acid (Asp) | 6.256 | 0.4 |
7. | Glutamic acid (Glu) | 6.696 | 0.4 |
8. | Threonine (Thr) | 7.027 | 0.4 |
9. | Glycine (Gly) | 7.466 | 0.4 |
10. | Alanine (Ala) | 8.227 | 0.4 |
11. | Proline (Pro) | 8.803 | 0.4 |
Peaks | tR (min) | Conc. (µmole/mL) | |
---|---|---|---|
12. | Methionine (Met) | 9.324 | 0.4 |
13. | Valine (Val) | 9.537 | 0.4 |
14. | Norvaline | 9.656 | 0.4 |
15. | Tryptophan (Trp) | 9.910 | 0.4 |
16. | Phenylalanine (Phe) | 10.120 | 0.4 |
17. | Isoleucine (Ile) | 10.214 | 0.4 |
18. | Leucine (Leu) | 10.285 | 0.4 |
19. | Cystine | 10.749 | 0.2 |
20. | Histidine (His) | 12.045 | 0.4 |
21. | Lysine (Lys) | 12.174 | 0.4 |
22. | Cysteine (Cys) | 12.463 | 0.4 |
23. | Tyrosine (Tyr) | 12.801 | 0.4 |
Column | Raptor ARC-18 (cat.# 9314A1E) | ||||||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Dimensions: | 100 mm x 3 mm ID | ||||||||||||||||||||||||||||||||
Particle Size: | 2.7 µm | ||||||||||||||||||||||||||||||||
Temp.: | 30 °C | ||||||||||||||||||||||||||||||||
Standard/Sample | |||||||||||||||||||||||||||||||||
Diluent: | 0.1 N HCl | ||||||||||||||||||||||||||||||||
Conc.: | 0.4 μmole/mL for each amino acid (0.2 μmole/mL for cystine) | ||||||||||||||||||||||||||||||||
Inj. Vol.: | 1 µL | ||||||||||||||||||||||||||||||||
Mobile Phase | |||||||||||||||||||||||||||||||||
A: | 0.1% Formic acid + 20 mM ammonium formate in water | ||||||||||||||||||||||||||||||||
B: | 0.1% Formic acid + 10 mM ammonium formate in 90:10 acetonitrile:water | ||||||||||||||||||||||||||||||||
|
Detector | UV/Vis @ 265, 4.8 nm |
---|---|
Instrument | Waters ACQUITY UPLC H-Class |
Sample Preparation | Derivatization reaction: 50 μL amino acid mix + 100 μL 0.2 N borate buffer (pH 10.0) + 50 μL 15 mM 9-fluorenylmethyl-chloroformate solution + 50 μL acetonitrile. The injection can be performed after 5 minutes of reaction time. |